TAVI - CoreValve (Medtronic): Difference between revisions

From Bay Area Structural Heart Wiki
mNo edit summary
 
(14 intermediate revisions by 3 users not shown)
Line 6: Line 6:
*[https://www.medtronic.com/us-en/healthcare-professionals/therapies-procedures/cardiovascular/transcatheter-aortic-valve-replacement.html CoreValve by Medtronic] - This page
*[https://www.medtronic.com/us-en/healthcare-professionals/therapies-procedures/cardiovascular/transcatheter-aortic-valve-replacement.html CoreValve by Medtronic] - This page
*[https://www.edwards.com/gb/devices/heart-valves/transcatheter-SAPIEN-3-Ultra Sapien 3 Ultra by Edwards] - [[TAVR - Sapien 3 (Edwards)|Sapien 3 procedure page]]
*[https://www.edwards.com/gb/devices/heart-valves/transcatheter-SAPIEN-3-Ultra Sapien 3 Ultra by Edwards] - [[TAVR - Sapien 3 (Edwards)|Sapien 3 procedure page]]
'''***Only use FX delivery systems, no longer using Pro+***'''


This procedure is usually done in Room 1, and requires a 5 person team: Front table scrub, Back table scrub (device prep), Circulator, Pacer/Circulator, and Recorder. The rep for the respective company will be present to assist with device prep except in the case of an emergency valve placement.
This procedure is usually done in Room 1, and requires a 5 person team: Front table scrub, Back table scrub (device prep), Circulator, Pacer/Circulator, and Recorder. The rep for the respective company will be present to assist with device prep except in the case of an emergency valve placement.
Line 24: Line 27:


*5Fr Micropunture  Kit
*5Fr Micropunture  Kit
*6Fr x 10cm sheath
*6Fr x 10cm sheath ( Not needed for valve-in-valve)
*8Fr x 10 cm sheath  
*8Fr x 10 cm sheath  
*8Fr x 35 cm Brite Tip Sheath
*SHEATHLESS INLINE DELIVERY: 18F coons dilator and 14F Cook sheath post. (Do not open until confirmed during timeout)
*SHEATHLESS INLINE DELIVERY: 18F coons dilator and 14F Cook sheath post.
** 18fr Dryseal if sheath delivery system is used or when doing a BAV.  
 
* 16fr Dryseal if sheath delivery system is used.


===Wires===
===Wires===
Line 35: Line 36:
*(2) Cordis 150J
*(2) Cordis 150J
*Cordis 260J
*Cordis 260J
*Amplatz SS 1cm tip
*Safari
*Terumo Stiff Straight Glidewire 0.035 x 150 cm
*Terumo Stiff Straight Glidewire 0.035 x 150 cm
*Amplatz 7cm tip if using dryseal  for sheath advancement
** .035 Standard Straight for Valve-in-Valve crossing, no need for glidewire.


===Catheters:===
===Catheters:===
Line 44: Line 47:
*5Fr Str Pigtail
*5Fr Str Pigtail
*6Fr JL4/JR4 (when doing coronary angiography)
*6Fr JL4/JR4 (when doing coronary angiography)
{{Warn|This has changed to 6 French!}}
*'''Valve-in-Valve cases: Mosaic valve needs pigtail, all others do not'''


===Medications on table:===
===Medications on table:===
Line 51: Line 54:


*Epinephrine 10mcg/ml in red medallion syringe (From anesthesia)
*Epinephrine 10mcg/ml in red medallion syringe (From anesthesia)
*BACK TABLE ;  2 L of cold saline for delivery catheter prep ; 1 L room temp Hep saline for front table ; 1 L room temp saline for back table 


===Miscellaneous Supplies===
===Miscellaneous Supplies===
*Acist hand control
*Acist hand control
*2L cold saline
*2L cold saline
*(3) Perclose
*(3) Perclose
*Spies 22G Grey Lido Needle
*Spies 22G Grey Lido Needle
* Dual Red/Black Pacing Remington cables
* EP Quad Cable


===Transvenous pacemaker supplies===
===Unipolar Transvenous pacing supplies.===
 
*Alligator Clip Remington Pacing Cables
*5Fr Josephson Pacing catheter for ''DD'' ; 5F Abbott Pacel Balloon Flow Directed Tip Pacing catheter for ''CS''
*TZ Medical Defibrillation Pads - Found in EP pyxis in labs. Large pad utilized as a grounding pad on the patients chest (subxyphoid)
*Dual Red/Black Pacing Remington cables
*NOTE : If patient is in second degree HB please prep RIJ possible
*EP Quad Cable
 
 
 
<br />


==='''PROCEDURAL INFO:'''===
==='''PROCEDURAL INFO:'''===
If a sheath is inserted prior to delivery system an 18F dryseal will be used.
*Contralateral arterial access - 6fr. Sheath (normally LFA) using micropuncture. Verify wire position with fluoro and save an image.
 
*Primary access - with micropuncture.  Verify wire position with fluoro and save an image
If delivery system is inserted sheathless, a 14F cook sheath will be inserted after removal of the delivery system.
**Pre-close x 1, sutures secured with hemostat.
**8F sheath reconstituted and advanced after pre-closure.


Straight pigtail is for root injections. 2nd pigtail is used to remove SS wire post-valve deployment
*Straight Pigtail -> LFA -> AO root injection
**Standard Injector settings (adjusted based on pt and root size)
'''5/5 on Acist'''
*AL1 advanced through 8F sheath over standard J.
*Standard J removed and crossing wire advanced. Cross valve. Wire removed.
*Measure EDP and gradient...maybe
*Working wire advanced to LV. AL1 removed.
*8F sheath removed and 18 F dilator in and out OR 16 F dryseal advanced depending if doing in-line or not.
*Check valve under fluoro for markers
*Insert valve delivery system
*Multiple AO grams for positioning
*Test pacing on Safarii wire
**Rapid pace 160, check for capture, 220 burst
*Pace -> AO gram -> Slowly deploy valve
**Cine length set at 40 seconds or more
*TTE
*Pigtail advanced to remove working wire from LV


Put stopcock on sideport of delivery system<br /><br />
*Pigtail readvanced to AO for AO gram (15/30 standard injection)
*Delivery system removed and 14F sheath advanced.
**Have second perclose, tools and dilator ready
*Protamine given, Preclosures completed, manual pressure for 5ish minutes
*Pigtail pulled back to iliac bifurcation and subtracted iliac angio performed (6/18 on Acist)
*Pigtail removed
*Secondary access Perclosed
*Dermabond all access sites, no dressing needed.<br /><br />
<hr />
<hr />
Approved: MM/YY
Approved: MM/YY
[[Category:Procedures]]
[[Category:Procedures]]

Latest revision as of 19:49, 19 January 2024

Transcatheter Aortic Valve Repair (Wikipedia) procedures utilize a femoral approach to deliver a stent-based aortic valve to the patients heart. There are two valve types available:

Error creating thumbnail: File with dimensions greater than 12.5 MP
Loading Instructions

File:Medtronic Valve Loading Video Small.mp4

***Only use FX delivery systems, no longer using Pro+***


This procedure is usually done in Room 1, and requires a 5 person team: Front table scrub, Back table scrub (device prep), Circulator, Pacer/Circulator, and Recorder. The rep for the respective company will be present to assist with device prep except in the case of an emergency valve placement.

In rare cases, the BASILICA procedure will be necessary to split the existing valve leaflets to prevent blockage of the coronary arteries. In that situation, the equipment on the BASILICA page will be needed in addition to the equipment listed below.

Per Rosalie:

TAVR/TMVI

  1. Edwards Sapien 3 Ultra (20, 23 & 26mm) and Sapien 3 (29mm) – scan the valve barcode (box) in McKesson for documentation and charges.  Bundle includes valve, delivery system, crimper, e sheath intro set, balloon cath & locking syringe.  With the new inventory system (POU) space is limited in entering item full description.   If pre dilatation Edwards balloon is necessary free text the balloon used in procedure notes (McKesson).
  2. Medtronic Core Valve – all items are charge separately.  Scan all barcodes (delivery system, loading system and valve).

Daniels/Spies - Draft

ACIST A1000V SYRINGE USED WITH 50/50 CONTRAST MIX

Sheaths:

  • 5Fr Micropunture Kit
  • 6Fr x 10cm sheath ( Not needed for valve-in-valve)
  • 8Fr x 10 cm sheath
  • SHEATHLESS INLINE DELIVERY: 18F coons dilator and 14F Cook sheath post. (Do not open until confirmed during timeout)
    • 18fr Dryseal if sheath delivery system is used or when doing a BAV.

Wires

  • (2) Cordis 150J
  • Cordis 260J
  • Safari
  • Terumo Stiff Straight Glidewire 0.035 x 150 cm
  • Amplatz 7cm tip if using dryseal for sheath advancement
    • .035 Standard Straight for Valve-in-Valve crossing, no need for glidewire.

Catheters:

  • 5Fr AL1 diagnostic
  • 5Fr Ang Pigtail
  • 5Fr Str Pigtail
  • 6Fr JL4/JR4 (when doing coronary angiography)
  • Valve-in-Valve cases: Mosaic valve needs pigtail, all others do not

Medications on table:

  • Lido/Bupivicaine 50/50
  • Epinephrine 10mcg/ml in red medallion syringe (From anesthesia)
  • BACK TABLE ; 2 L of cold saline for delivery catheter prep ; 1 L room temp Hep saline for front table ; 1 L room temp saline for back table

Miscellaneous Supplies

  • Acist hand control
  • 2L cold saline
  • (3) Perclose
  • Spies 22G Grey Lido Needle
  • Dual Red/Black Pacing Remington cables
  • EP Quad Cable

Unipolar Transvenous pacing supplies.

  • Alligator Clip Remington Pacing Cables
  • TZ Medical Defibrillation Pads - Found in EP pyxis in labs. Large pad utilized as a grounding pad on the patients chest (subxyphoid)
  • NOTE : If patient is in second degree HB please prep RIJ possible

PROCEDURAL INFO:

  • Contralateral arterial access - 6fr. Sheath (normally LFA) using micropuncture. Verify wire position with fluoro and save an image.
  • Primary access - with micropuncture. Verify wire position with fluoro and save an image
    • Pre-close x 1, sutures secured with hemostat.
    • 8F sheath reconstituted and advanced after pre-closure.
  • Straight Pigtail -> LFA -> AO root injection
    • Standard Injector settings (adjusted based on pt and root size)

5/5 on Acist

  • AL1 advanced through 8F sheath over standard J.
  • Standard J removed and crossing wire advanced. Cross valve. Wire removed.
  • Measure EDP and gradient...maybe
  • Working wire advanced to LV. AL1 removed.
  • 8F sheath removed and 18 F dilator in and out OR 16 F dryseal advanced depending if doing in-line or not.
  • Check valve under fluoro for markers
  • Insert valve delivery system
  • Multiple AO grams for positioning
  • Test pacing on Safarii wire
    • Rapid pace 160, check for capture, 220 burst
  • Pace -> AO gram -> Slowly deploy valve
    • Cine length set at 40 seconds or more
  • TTE
  • Pigtail advanced to remove working wire from LV
  • Pigtail readvanced to AO for AO gram (15/30 standard injection)
  • Delivery system removed and 14F sheath advanced.
    • Have second perclose, tools and dilator ready
  • Protamine given, Preclosures completed, manual pressure for 5ish minutes
  • Pigtail pulled back to iliac bifurcation and subtracted iliac angio performed (6/18 on Acist)
  • Pigtail removed
  • Secondary access Perclosed
  • Dermabond all access sites, no dressing needed.


Approved: MM/YY