BATMAN
From Bay Area Structural Heart Wiki
BATMAN and UNICORN
- 18Fr Dryseal for non-inline Evolute (Confirm before opening)
- Amplatz Extra Stiff for Sheath advancement
- Suture for dryseal securement
- Micropuncture
- 8F sheath
- (2) .035 x 150 J wires
- .035 x 260 J wire
- Terumo Stiff Straight Glidewire 0.035 x 150 cm
- (2) Safari Wire
- (3) Perclose
- 7Fr MP1 Guide
- 5Fr AL1 diagnostic
- (2) 5Fr Ang Pigtail
- .014 Astato xs 20 guidewire 300cm
- FineCross 135cm
- Bovie pen
- Grounding Pad
- Bovie machine
- (2) Watchdog devices
- (2) Stopcocks
- (2) Needle drivers
- Alligator pacing wire
- Navicross
Standby - PULLED and HANGING on WALL
- Tip deflecting wire
- Sentinel
Additional Standby Items - don't need to have pulled
- MPA1, JR4, IM, AR1 (Right leaflet)
Procedure
- Setup and access is the same as TAVR except a 18F Dryseal sheath is inserted initially.
- Bovie set at 30 CUT ONLY.
- AL1 advanced over J wire through 7F sheath. Wire exchanged for straight stiff glidewire for valve crossing.
- AL1 exchanged over the wire for MP1 Guide with watchdog and stopcock attached.
- Platinum Plus wire is preshaped by MD and inserted through MP to retain LV access.
- Snare advanced through MP alongside Platinum Plus.
- BASILICA catheter advanced through Dryseal sheath with watchdog and stopcock attached.
- Kinked Astato wire advanced through BASILICA catheter.
- Hand injection performed to verify position of BASILICA catheter.
- Astato wire connected to bovie with 2 hemostats.
- RF applied as Astato wire advanced.
- Once wire penetrates leaflet it is snared and pulled into MP1.
- Platinum Plus wire removed
- AL1 advanced next to BASILICA catheter and valve recrossed in standard fashion.
- 60ml D5 filled syringes and extension tubing attached to sideports of guides and injected simultaneously during leaflet laceration.
- MP removed. BASILICA catheter removed.
- Amplatz ES advanced through AL1 and AL1 removed.
- Pigtail advanced to coronary cusp
- Dryseal exchanged for ESheath
- Do a TAVR.