Duo + Octaray/Optrell Lab 5 Pinning

From Bay Area Structural Heart Wiki
Revision as of 21:56, 27 February 2025 by Rosenbzj (talk | contribs) (Created page with "If Drs. Goyal/Tooley want to simultaneously use an Abbott DuoDeca and an Octaray or Optrell in lab 5 we need to change where the duodeca gets pinned. After selecting the appropriate Pruka protocol: Plug a 20-pole boat into Ref Deca and put the following pins into ports 1 through 10 (left side): (1,2,,5,6,,9,10,,11,12,,19,20). Please note, 11,12 and 19,20 are pins 1,2 and 9,10 on the black cable. Plug the following poles into ports 21-30 on the teal accessory block (co...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

If Drs. Goyal/Tooley want to simultaneously use an Abbott DuoDeca and an Octaray or Optrell in lab 5 we need to change where the duodeca gets pinned. After selecting the appropriate Pruka protocol:

Plug a 20-pole boat into Ref Deca and put the following pins into ports 1 through 10 (left side): (1,2,,5,6,,9,10,,11,12,,19,20). Please note, 11,12 and 19,20 are pins 1,2 and 9,10 on the black cable.

Plug the following poles into ports 21-30 on the teal accessory block (connected to block A on the front of the Pruka Amp) (3,4,,7,8,,13,14,,15,16) Please note, 13,14 and 15,16 are pins 3,4 and 5,6 on the black cable.

Select the appropriate