Tricuspid ViV
From Bay Area Structural Heart Wiki
Anesthesia | Imaging | Access | ACIST | Manifold |
---|---|---|---|---|
MAC | Fluoro, TTE |
Femoral Vein |
No | No |
Supplies
- Standard TAVR backtable setup
- 8F Sheath
- 150 J
- 260J
- Safari
- Alligator Clip Pacing cable
- 6F Angled Pigtail
- 0 Prolene
- Micropuncture (for woggle only)
- Perclose and Nu-Knit
Standby:
- 8.5F Agilis or TourGuide
- Double-pump-valve-cracker setup for BVF
- 6F balloon wedge
- 5F balloon tip pacer
Pacing - 25 Output, 140-200bpm
Procedure
- Single Venous access
- Pre-close
- 8F Sheath inserted
- Pigtail and J wire advanced to RV
- J wire exchanged for Safari
- Shave back of Safari for pacing
- Sheath exchanged for E Sheath
- Valve delivery system advanced
- Pacer tested
- Valve deployed
- Safari exchanged for J wire
- Perclose completed and woggle placed.
- Voila.