Setting up Abbott

From Bay Area Structural Heart Wiki
Revision as of 17:59, 16 August 2022 by Rosenbzj (talk | contribs) (Created page with "'''I.) St Jude Splitter (aka: Squid) -> GE Amp''' A) Cable A from GE Amp -> Squid Port A ; Squid A Cable-> GE Amp Port A B) ECG Cable from GE Amp -> Squid Port ECG ; ECG Cable...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

I.) St Jude Splitter (aka: Squid) -> GE Amp A) Cable A from GE Amp -> Squid Port A ; Squid A Cable-> GE Amp Port A B) ECG Cable from GE Amp -> Squid Port ECG ; ECG Cable from Squid -> GE Amp ECG Port C) Pin Box D Cable (St Jude Cart Drawer)-> Squid Port D; Squid Cable D -> GE Amp Port D

 1)	Plug in EGM Cable (St Jude Cart Drawer): EGM port on Ampere (green) -> Pin Box D 29-32

II.) Tactisys A) Black Box Power Cable on top of St Jude Cart (black/grey) -> Tactisys B) Generator Cable Ampere (yellow/ in cart drawer) -> Tactisys RF Port (green)

III.) St Jude Mapping Station Cables A) Ethernet Cable -> Ethernet Port (back of Tactisys) B) Amplifier Fiberoptic Cable Set -> Amplifier Ports (Amplifier on St Jude Cart) C) Ampere Fiberoptic Cable Set -> Ampere Ports (Ampere on St Jude Cart) D) Fiberoptic Cable (grey/black ONLY) -> Ultrasound Port (on light wave box for Vantageview)

IV.) St Jude Control Room Cables A) GE VGA Cable -> GE VGA Port (back of Ampere) B) RF Remote Cables -> RF Remote Cable Ports (back of Ampere)

V.) Ablation Pump A) Power Cable (St Jude Cart drawer) B) VGA Pump Cable (back of St Jude Cart) -> VGA Port (back of Cool Point Pump)

VI.) Patient Patching A) Secure Magnet to table with connected Magnet Cable (Field Port on side of Status-> Magnet) B) NavLink (St Jude Drawer): place on table and connect to St Jude Green Port on Amplifier C) PRS-A and PRS-B Cables (aka: magnet cables): place on patient table (Connected to Ref Ports on front of Status)

VII.) Power up System A) Plug in: Power Cable on Back of Power Box (St Jude Cart) to grounded outlet B) Turn on Power Switches: Power Box, Amplifier, Ampere, Tactisys, Ablation Pump

VIII.) Control Room A) Turn Black Knob (base of electrical tower): Carto B -> Abbott A B) CardioLab: Stockert -> Null -> Ibi1500T (study configuration, ablation, drop down menu select Ibi1500T, Apply, Close) At the end of the case turn Ibi1500T -> Null C) VantageView: Woods