Note Note: Apply Foresight to patient before draping to save yourself the hassle later.

Daniels/Spies - Draft

Also used for Dr. Sheridan

Wires:

Open

  • Amplatz SS 7cm tip (x 2 for Spies)
  • Cordis 150J

Standby

  • Amplatz Extra Stiff
  • Glidewire

Sheaths:

Open

  • 8fr Pinnacle

ECMO cart:

Open

  • ECMO instruments
  • 16/18 Coons dilators
  • 20/22 Coons Dilators
  • Arterial vented cannula - Normally 16fr.
  • Venous cannula - Normally 21fr.
  • Irrigation bulb syringe
  • ”Carrot tops” Orange vessel tourniquets

Standby

  • Alternative sizes of cannulas
  • Arrow 5 and 6fr braided sheaths for antegrade bypass
  • Male Male luer adapter for antegrade bypass


Misc Supplies:

Open

  • Micropuncture
  • Probe cover
  • Perclose
  • (2) 0 Ethibond (Not for Sheridan)
  • (2) 2-0 Prolene (Open 4 for Dr. Sheridan)
  • (1) White Nylon
  • (2) Biopatch

* Be prepared to perform a septostomy post ECMO implant *

Egrie

  • Substitute Vascular dilator kit for Coons dilators
  • Substitute white nylon and Martin free needles for securement suture

Procedure

  1. Table setup:

Note Note: Spies would like bi-fem, right radial and right IJ prepped. Patient will need a leave-in swan and RIGHT radial A-line before leaving room.

Note Protip:
Separate Coons dilators. We use the 16 and 20 French. The other can go to the back table for standby. This eliminates confusion during the rapid access portion of the procedure.
Note Protip:
Arrange instruments. Have 6 tubing clamps and heavy trauma shears immediately available to the physician. Needle driver and Mayo scissors are used for suturing after cannula are in place.

  1. Arterial access with ultrasound and micro puncture (Usually left groin)
  2. Short sheath wire and 8fr. dilator
  3. Pre-close
  4. Stiff wire inserted
  5. 16fr. Coons dilator
  6. Arterial (Aortic) Cannula inserted
  7. Pull wire and dilator while MD clamps
  8. Venous access (Daniels uses big needle and ultrasound)(Usually right groin)
  9. Stiff wire inserted (Usually no pre-close)
  10. 20fr. Coons dilator
  11. Venous cannula inserted (Hold dilator in place in cannula while advancing)
  12. Pull wire and dilator while MD clamps
  13. CST hands over sterile portion of circuit
  14. Clamp circuit while MD cuts tubing
  15. Assist MD in cutting and untwisting circuit
  16. Use irrigation syringe to assist MD in air-free connection between circuit and cannula
  17. 0-Prolene used to create purse-string closures around each cannula
  18. 0-Ethibond used to secure cannulas to skin
  19. Suture-locks used ~12” down tubing to further secure circuit
  20. Carrot-top cut to ~5” used to protect Perclose sutures from being pulled or cut
  21. Bio-patches placed
  22. Clean and dress both sites

Securement and anti-bleeding protocol

At the Skin:

  • 0-Prolene purse-strings on both cannula secured with Woggle devices.
  • Additional deep 0-Prolene stitch through a 3” long carrot-top secured to the cannula with white nylon ties. (Minimum 2 nylon ties)
  • Arterial cannula secured with 0-Ethibond at the skin in the provided suture ring.

At the tubing connection

  • 0-Ethibond deep skin sutures double-wrapped and tight around the cannula. Both sides.

~8-10”down

  • 0-Ethibond deep skin sutures double-wrapped and tight around the cannula. Both sides.
  • Suture locks can be used above the knees for further strain relief.

Note Note: Surgicel can be used at the skin for further bleeding control.

Septostomy

See Septostomy page

Antegrade Perfusion

  • Use Arrow short braided sheath on ECMO cart
  • Ultrasound and micro puncture used to access SFA below arterial cannula
  • Use double-male lure adaptor to attach sheath side-port to arterial cannula vent.

Spies' Decan without Pre-Closure!

  • 12F Dilator
  • 10F Dilator
  • 8F Dilator
  • 2 Perclose
  • Standard J

Clamp arterial cannula, cook needle stick to the cannula. Standard J advanced and cannula removed. 12F dilator inserted, pressure held for 5 or so minutes. Dilator downsized to 10F, pressure held. Perclose times one. 8F dilator advanced and pressure held. Second perclose deployed.


Approved: MM/DD