Micra post TAVI

From Bay Area Structural Heart Wiki
Revision as of 21:53, 5 August 2021 by Suddemj (talk | contribs) (Created page with "In the course of a TAVI, it is sometimes necessary to implant a leadless pacemaker. This facility uses the Micra device. This page will guide you on this procedure in the post...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

In the course of a TAVI, it is sometimes necessary to implant a leadless pacemaker. This facility uses the Micra device. This page will guide you on this procedure in the post-TAVI setting.

Already on the table:

  • Micropuncture set
  • 150 j-wire
  • 2x 20cc syringes
  • 1-2x 60cc syringes

Open:

  • 14 and 22 coons dilators
  • 0 PDS or Prolene
  • 180 Amplatz Super Stiff, 7cm tip

If pt. is dependent on TPM and it is in the RFV, assume you will be switching the TPM to the LFV and open:

  • New ultrasound probe cover
  • 8.5 x 35 Brite-tip sheath
  • 5fr. Ballon tip TPM catheter

Prep:

  • Set ACIST at 5ml/sec, 20ml total (to fill syringes).