TMVI
From Bay Area Structural Heart Wiki
Warning: This page needs revision! TransCatheter Mitral Valve Repair/Replacement
Per Rosalie:
TAVR/TMVI
- Edwards Sapien 3 Ultra (20, 23 & 26mm) and Sapien 3 (29mm) – scan the valve barcode (box) in McKesson for documentation and charges. Bundle includes valve, delivery system, crimper, e sheath intro set, balloon cath & locking syringe. With the new inventory system (POU) space is limited in entering item full description. If pre dilatation Edwards balloon is necessary free text the balloon used in procedure notes (McKesson).
- Medtronic Core Valve – all items are charge separately. Scan all barcodes (delivery system, loading system and valve).
MITRAL CLIP (Mitral/Tricuspid):
- Use the barcode on Mckesson screen in Lab 1 – Mitraclip (Abbott) bundle consists of (1) Steerable Guide Catheter, (2) NTR Clip Delivery System & (1) XTR Clip Delivery System.
GENERAL ANESTHESIA AND TEE IMAGING, RADIAL ART LINE BY ANESTHESIA
ACCESS 2 FEMORAL VEINS (TAVR AND PACER)
PRE PROCEDURE: TYPE AND CROSS, ASPIRIN GIVEN
SHEATHS:
- Micropuncture
- 8fr Pinnacle for Perclose
- 8x35 Brite tip
- 8.5 SL1
- 8.5 Tourguide
WIRES:
Open
- 150 J wire
- .032 x 260 J-wire
- Safari 2
- Amplatz Super Stiff 1cm tip
CATHETERS
Open
- 6fr. Straight Pig
- 5fr. MPA2
- Naked 3 Port Manifold
PACING
- 5F Josephson Pacing Cable
- EP Quad Cable
- Remington Pacing Cable
- (2) Pacer Boxes
FOR EDWARDS VALVE
- INDEFLATOR
- CRIMPER
- E SHEATH
- DELIVERY SYSTEM
- VALVE
TRANS SEPTAL
- BAYLIS NEEDLE LARGE CURVE C1
- BAYLIS MACHINE
- BAYLIS CABLE
- GROUNDING PAD
SEPTAL DILATION
Open
- 12X4X75 MUSTANG OR 14X4 XXL (Ask before opening)
- PRESTO INDEFLATOR
Standby
- 9X4X75 MUSTANG
- 14, 16 XXL Balloons
- TORQUEVUE 45DEGREE DELIVERY SYSTEM (9F OR 10F)
- 10MM, 14MM, 18MM SEPTAL OCCLUDER
ADDITIONAL SUPPLIES
Open
- 2xPERCLOSE
- 3 PORT MANIFOLD - No contrast
- PROBE COVER
- PROLENE 2-0
- FLUORO DRAPE
- CHLORAPREP AND PREP TRAY
Standby
- 150ML CONTRAST (FOR CORONARIES)
- BIOPATCH
- 8F ANGIOSEAL
- 0 SILK SUTURE
- EXTRA PERCLOSE
MEDICATIONS
- BUPIVACAINE
- EPINEPHRINE 10MCG/ML IN RED SYRINGE ON BACK TABLE
- ROTAGLIDE 10ML (EDWARDS VALVE)
STANDBY EMERGENCY ECMO SUPPLIES
- 16F ARTERIAL CANNULA
- 21F VENOUS CANNULA
- (2) PERCLOSE
- ECMO INSTRUMENT TRAY
- AMPLATZ SUPER STIFF
STANDBY PERICARDIOCENTESIS SUPPLIES
- EYE DRAPE
- CHIBA NEEDLE 10CM 18G
- SKATER INTRODUCER SET
- BARD LOCKING PIGTAIL 6F OR 8F
- (3) 60ML SYRINGES
- STOPCOCK
- AMPLATZ SUPER STIFF 180CM
PROCEDURE:
- Patient prepped from navel to knees in sterile fashion.
- Access obtained with u/s
Typical sites:
14F OR 16F RFV E-sheath for valve delivery
- Dilate with 8F
- Preclose
- SL1 inserted over .032 260cm J
- Baylis needle inserted through SL1 and atrial septum puncture performed.
- Needle removed, .032 wire advanced to pulmonary vein
- MP advanced to LUPV, .032 wire removed and super stiff advanced
- MP removed and agilis advanced and directed to mitral valve
- Pigtail inserted through agilis for valve crossing.
- Safari2 wire inserted to ventricle.
- Pigtail and agilis removed.
- 12mm/14mm balloon advanced for septostomy and possible BAV
- Balloon removed, valve delivery system advanced.
- When valve is in place, temp pacer inserted and tested.
- Valve deployed. Septum evaluated with echo for shunting.
- Pacer removed. Delivery system removed.
- Preclosure closed and stitch to RFV
- 8F vein pulled and manual pressure applied
- Tegaderm both sides.
APPROVED: MD initials MM/YY