Quantien RFR wire (Abbott)

From Bay Area Structural Heart Wiki
Revision as of 20:56, 9 June 2020 by Suddemj (talk | contribs) (Created page with "This is the iFR and FFR wire currently used in the cath lab. #Log into the device username: cpmcvn, password:cmpcvn #If device was powered off less than 5 minutes ago, it w...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

This is the iFR and FFR wire currently used in the cath lab.

  1. Log into the device username: cpmcvn, password:cmpcvn
  2. If device was powered off less than 5 minutes ago, it will ask if you want to continue the same patient.
  3. Select the room you are in.
  4. Enter patient info. (Can be entered at the end for STEMI)
  5. Follow on-screen prompts
  6. Re-zero PA pressure on device and (Optional)McKesson
  7. Flush wire and place flat on table
  8. Press connect on machine, then turn on wire transmitter
    • Light should be solid green if connected
  9. Introduce wire into guide and advance until radio-lucent portion is outside the guide
  10. Equalize
  11. Select mode at top, center of screen
    • RFR - Resting full-cycle Ratio (AKA - IFR)
      • Average of 5-beat difference through full cardiac cycle
      • Camera icon gives a single reading
      • Pullback button activates pullback
    • FFR - Traditional measurement using Adenosine