Woods Farapulse
last edited 9/5/24 zr/jc
work in progress
General:
-Consent - Electrophysiology Study with Catheter Ablation
-Patient intubated/paralyzed (long acting) throughout procedure
-No foley
-No isuprel
-Antibiotics per woods protocol (only w/device, implants etc.)
-Fluoro
-Goal ACT>300
-MacLab leads, Anesthesia leads, Claris 12 lead
-2 Grounds, Abbott patches, regular defib pad
From Pyxis:
-Case by case basis for protamine
-No famotidine
-bupivicaine
- (2x) heparin vials (10,000 u/10mL)
- (1x) heparin premixed (25,000/250mL)
- for typical AFL PFA: Nitro - 250ml bottle of 200mcg/ml
Bags:
-500mL NS for ACT line
-1L LR anesthesia
-(1x) 500mL NS with 5000u heparin for the table
-(2x) 500mL NS with 5000u heparin for red/yellow lines
-(1x prepulled from pyxis 25,000u heparin/250mL NS) heparin drip for blue line
Lines:
-ACT
-Blue (heparin drip)
-Red (Grid - Heparin 5000u/500mL NS @ 120cc/hr)
-Yellow (Farapulse - Heparin 5000u/500mL NS pressure bag, no pump)
Charting ablation in Cupid:
System used - freetext Farapulse
Energy - freetext PFA
Number of energy applications - # per Ablation site
Drop List:
Sheaths:
8F pinnacle (ICE)
8f locking sheath (inquiry unless expectation of CTI ablation - in which case DuoDeca)
8f pinnacle (---> Preclose ----> SL0 + Baylis 71cm ---> Map with Grid -----> Remove Grid/Upsize to Faradrive ----> Ablate with Farawave)
Rampart + Radpad
3x yellow filters
8f Acunav (non-soundstar)
1x Perclose
Inquiry + Cable (unless CTI - then DuoDeca + Cable)
Grid + Cable
short versacross kit - 180cm pigwire
Light handle cover
Alligator Clips
SL0/Baylis 71cm (from lab5)
1x Ice Sleeve
1x Access sleeve
ACT line
3x pressure lines (label blue/red/yellow)
Ioban
8F pinnacle x2
8f locking sheath
ask - driver/0pds/stopcock for woggle if no protamine
chloraprep
vascade x2
Rosen wire w/1.5mmJ
created 5/7/24 ac/zr