Tricuspid ViV

From Bay Area Structural Heart Wiki
Revision as of 22:25, 8 December 2021 by Suddemj (talk | contribs)
Anesthesia Imaging Access ACIST Manifold
MAC Fluoro, TTE

Femoral Vein

No No

Supplies

  • Standard TAVR backtable setup
  • 8F Sheath
  • 150 J
  • 260J
  • Safari
  • Alligator Clip Pacing cable
  • 6F Angled Pigtail
  • 0 Prolene
  • Micropuncture (for woggle only)

Standby:

  • 8.5F TourGuide
  • 6F balloon wedge
  • 5F balloon tip pacer

Procedure

  1. Single Venous access
  2. Double Preclose
  3. 8F Sheath inserted
  4. Pigtail and J wire advanced to RV
  5. J wire exchanged for Safari
    • Shave back of Safari for pacing
  6. Sheath exchanged for E Sheath
  7. Valve delivery system advanced
  8. Pacer tested
  9. Valve deployed
  10. Safari exchanged for J wire
  11. Perclose completed and woggle placed.
  12. Voila.