Carotid TAVI
From Bay Area Structural Heart Wiki
Procedure to be performed in Hybrid OR 14 in conjunction with the OR team.
Carotid TAVR | ||
---|---|---|
Anesthesia | Imaging | Access |
General | Fluoro and TEE |
|
Pre-Procedure Watchouts | ||
*Transcutaneous O2 Sat
|
Equipment
In addition to standard TAVR equipment the OR team will need:
- Micropuncture
- 6F 10cm Pinnacle Sheath
We will need:
- Safari wire instead of Amplatz ES
- Can get away with 1 short J wire
Setup- Right Sided Carotid
- Anesthesia to get left sided A line
- Table at 30 degrees
- Bring C arm in perpendicular to table
- Echo at head of bed (TEE)
- OR will do prep. (Include groins and right wrist)
- First assist to stand at head of bed on the left
- CT surgeon on right side with scrub RN
- Cath Lab will have standard 2 tables plus high table for valve delivery.
- mayo stand to bridge the gap between valve table and access site
- Cover Fluoro Pedal and Base of C-arm
- 2 venous extensions needed for Anesthesia
- Leave large boom to the side of the room. Use Monitors 3 &4 for pigtail and pacer.
- 3 Live and 4 McKesson
- When DD moves to the head to assist with valve crossing bring in monitors 3 &4
- 3- echo 4-Mckesson 1- Live 2- Anesthesia Vitals
- Right wall monitor with quad view
- boom with standard tavr display
Procedure- Right Sided
- Sheridan will start with cutdown.
- DD will obtain radial access for pigtail and femoral venous for Pacer
- AOgram per usual
- CTS inserts 6F sheath
- One CL tech stays at foot of bed for injections and pigtail pull
- Preloaded AL1 and Stiff glide inserted
- stiff glide exchanged for SAFARI
- E sheath inserted
- Valve delivered
- Venous sheath removed and figure 8 with prolene
- TR Band applied after OR finished closing.