The Cardiomems device is used to passively monitor Pulmonary Artery pressures.
The Cardiomems device is used to passively monitor Pulmonary Artery pressures.
Procedure type in McKesson: (33289) IMPL WRLS PA PRESSURE SENSOR




==Dr. Herr==
==Dr. Herr==
===Open===
===Open===
*US probe cover
*US probe cover
*Micropuncture
*Micropuncture
Line 17: Line 21:


===Standby===
===Standby===
*Cardiomems Device
*Cardiomems Device




===Procedure===
===Procedure===
*No need for contrast manifold, Dr. Herr prefers to inject with a syringe
*No need for contrast manifold, Dr. Herr prefers to inject with a syringe
*75 Amplatz is in IR
*75 Amplatz is in IR
Line 31: Line 37:
#Swan inserted for full Right Heart Cath (.025 wire used if needed to get to PA)
#Swan inserted for full Right Heart Cath (.025 wire used if needed to get to PA)
#Dr. Herr prefers Left PA for the device
#Dr. Herr prefers Left PA for the device
#Pulmonary Arteriogram for anatomy and placement planning(Glowtape provides reference point)
#Pulmonary Arteriogram for anatomy and placement planning (Glowtape provides reference point)
#Open device, flush wire lumen, and whisk in saline for 10-30 seconds
#Open device, flush wire lumen, and whisk in saline for 10-30 seconds
#Insert device and advance to PA
#Insert device and advance to PA

Revision as of 16:25, 17 December 2019

The Cardiomems device is used to passively monitor Pulmonary Artery pressures.


Procedure type in McKesson: (33289) IMPL WRLS PA PRESSURE SENSOR


Dr. Herr

Open

  • US probe cover
  • Micropuncture
  • 75cm or 180 cm Amplatz Super Stiff
  • 7fr. And 9fr. dilators
  • 12fr. FastCath sheath
  • 7fr. TD swan
  • .025 wire
  • Steelcore
  • 2-port Pressure manifold
  • ~20cc contrast
  • Glowtape (Usually behind room 6)

Standby

  • Cardiomems Device


Procedure

  • No need for contrast manifold, Dr. Herr prefers to inject with a syringe
  • 75 Amplatz is in IR
  1. Place Glowtape under patient, slightly left of spine, mid chest
  2. Right groin access with micropuncture and ultrasound
  3. Amplatz wire inserted in micropuncture sheath
  4. 7 and 9 French dilators
  5. 12 French Fast Cath
  6. Swan inserted for full Right Heart Cath (.025 wire used if needed to get to PA)
  7. Dr. Herr prefers Left PA for the device
  8. Pulmonary Arteriogram for anatomy and placement planning (Glowtape provides reference point)
  9. Open device, flush wire lumen, and whisk in saline for 10-30 seconds
  10. Insert device and advance to PA
  11. Device is deployed, delivery system removed, Steelcore pulled back to common PA
  12. Swan advanced over steelcore to right PA for calibration numbers
  13. Cardiomems calibrated
  14. Sheath pulled manual pressure only.