LAAO with Amulet: Difference between revisions

From Bay Area Structural Heart Wiki
Warneh (talk | contribs)
No edit summary
Warneh (talk | contribs)
No edit summary
Line 37: Line 37:
* Wholey / Versacore
* Wholey / Versacore


=== Amulet supplies: ===
=== Amulet supplies: (supplies provided by Rep.) ===


* 12 Fr or 14 Fr Tourque View Delivery (Open as requested after TEE)
* 12 Fr or 14 Fr Tourque View Delivery (Open as requested after TEE)
Line 51: Line 51:
# Baby J wire from sheath
# Baby J wire from sheath
# 8 Fr dilator
# 8 Fr dilator
# 6 Fr Perclose pre-close RFV.  .032 wire inserted after preclosure deployment.
# 6 Fr Perclose pre-close RFV.  
# .035 J wire from versacross system advanced thru perclose for Baylis sheath advancement.
# .035 J wire from versacross system advanced thru perclose for Baylis sheath advancement.
# .035 J wire removed, VersaCross RF pigtail wire advanced for transeptal puncture.
# .035 J wire removed, VersaCross RF pigtail wire advanced for transeptal puncture.
Line 58: Line 58:
# Transeptal puncture, ACT drawn from LA.
# Transeptal puncture, ACT drawn from LA.
# Measure LA pressure
# Measure LA pressure
# Tourque View Delivery advanced to the LA
# ? Tourque View Delivery advanced to the LA
# Backloaded 5 Fr Str Pigtail advanced to the LAA
# ? Backloaded 5 Fr Str Pigtail advanced to the LAA
# Hook pigtail up to manifold for angiogram of LAA
# ? Hook pigtail up to manifold for angiogram of LAA
# Measure and select size of device
# Measure and select size of device
# Open and prep the device
# Open and prep the device

Revision as of 16:51, 29 May 2025

Pre-Procedure Watchouts: Type and Screen
Anesthesia Imaging Access Acist/Manifold
Gernerl Fluoro and TEE (ICE?) 12 Fr or 14Fr RFV Manifold with contrast
Rep: Monica Tseng 415-816-2877

Open

  • Micropuncture
  • 8fr sheath (dilator used pre-Perclose)
  • Probe cover
  • 1 Perclose
    • 0-Prolene or 0-PDS (Spies will add a suture after perclose if site is oozing. If no ooze Exofin.)
  • Cordis 150J
  • 5fr Straight Pigtail
  • 3-port manifold
  • Baylis VersaCross Access System (purple box in structural cart)
  • Grounding pad
  • (Spies - Lido with EPI)

Standby

  • 4 or 5fr Pinnacle for Arterial line
  • 190 Supracore
  • Wholey / Versacore

Amulet supplies: (supplies provided by Rep.)

  • 12 Fr or 14 Fr Tourque View Delivery (Open as requested after TEE)
  • Amulet device (Size as requested after TEE)

PROCEDURE

  1. Basic table set up ONLY until it is determined by TEE that case will proceed.
  2. TEE performed. Procedure aborted if clot
  3. Prep bilateral groins and drape once probe has been inserted.
  4. Micropuncture ultrasound-guided RFV access
  5. 5 Fr Micropuncture sheath
  6. Baby J wire from sheath
  7. 8 Fr dilator
  8. 6 Fr Perclose pre-close RFV.
  9. .035 J wire from versacross system advanced thru perclose for Baylis sheath advancement.
  10. .035 J wire removed, VersaCross RF pigtail wire advanced for transeptal puncture.
  11. TEE & fluoro guided transeptal access & images of Left Atrial Appendage
  12. Anesthesia gives Heparin IV. ACT GOAL >250-300
  13. Transeptal puncture, ACT drawn from LA.
  14. Measure LA pressure
  15. ? Tourque View Delivery advanced to the LA
  16. ? Backloaded 5 Fr Str Pigtail advanced to the LAA
  17. ? Hook pigtail up to manifold for angiogram of LAA
  18. Measure and select size of device
  19. Open and prep the device
  20. Pigtail removed just prior to delivery system insertion.
  21. Placement confirmed by LA angiogram
  22. Delivery system removed as MD tamps perclose knot.
  23. If access is clean, Dermabond. If bleeding, woggle