Maclab FFR/DFR: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
|||
Line 1: | Line 1: | ||
__NOTOC__ | |||
==Step 1== | ==Step 1== | ||
*Turn on P2 channel by either double clicking P2, or clicking P2 on keyboard | *Turn on P2 channel by either double clicking P2, or clicking P2 on keyboard |
Latest revision as of 21:05, 26 February 2025
Step 1
- Turn on P2 channel by either double clicking P2, or clicking P2 on keyboard

Step 2
- Open REAL TIME tab, then select FFR display on. This step will not work if you have not opened P2 yet.

Step 3
- Once wire is inserted and physician is ready, equalize by double clicking in FFR square.
- After equalized switch to DFR if not using adenosine.
THIS MAY TAKE A FEW ATTEMPTS FOR PRESSURE TO EQUALIZE, THIS IS NORMAL!!!!!
Step 4
- Once equalized and on desired measurement selection (FFR or DFR), open MEASUREMENTS tab and select FFR/DFR

Step 5
- A window will open on the RIGHT sided monitor, all you have to do is click measure, you do NOT have to select a vessel.
- The measurement should populate under baseline measurements after a few beats.