Step 1 -
==Step 1==
      Turn on P2 channel by either double clicking P2, or clicking P2 on keyboard
*Turn on P2 channel by either double clicking P2, or clicking P2 on keyboard
[[File:FFR Step 1.jpg|thumb|218x218px]]
[[File:FFR Step 1.jpg|left|thumb|218x218px]]


<div style="clear: both"></div>
==Step 2==
*Open REAL TIME tab, then select FFR display on. This step will not work if you have not opened P2 yet.


[[File:FFR step 2.jpg|left|thumb|223x223px]]


<div style="clear: both"></div>


==Step 3==
*Once wire is inserted and physician is ready, equalize by double clicking in FFR square.
*After equalized switch to DFR if not using adenosine.
'''THIS MAY TAKE A FEW ATTEMPTS FOR PRESSURE TO EQUALIZE, THIS IS NORMAL!!!!!'''




 
==Step 4==
Step 2 -
*Once equalized and on desired measurement selection (FFR or DFR), open MEASUREMENTS tab and select FFR/DFR
 
Open REAL TIME tab, then select FFR display on. This step will not work if you have not opened P2 yet.
 
[[File:FFR step 2.jpg|thumb|223x223px]]
 
 
 
 
 
 
 
 
 
Step 3 -
 
Once wire is inserted and physician is ready, equalize by double clicking in FFR square. After equalized switch to DFR if not using adenosine. '''THIS MAY TAKE A FEW ATTEMPTS FOR PRESSURE TO EQUALIZE, THIS IS NORMAL!!!!!'''
 
 
Step 4 -
 
Once equalized and on desired measurement selection (FFR or DFR), open MEASUREMENTS tab and select FFR/DFR
[[File:FFR Step 3.jpg|thumb|269x269px]]
[[File:FFR Step 3.jpg|thumb|269x269px]]




<div style="clear: both"></div>


==Step 5==
*A window will open on the RIGHT sided monitor, all you have to do is click measure, you do NOT have to select a vessel.
*The measurement should populate under baseline measurements after a few beats.   






Step 5 -
A window will open on the RIGHT sided monitor, all you have to do is click measure, you do NOT have to select a vessel. The measurement should populate under baseline measurements after a few beats.   
<nowiki>[[Category:Setups]]</nowiki>


[[Category:Setups]]
[[Category:Setups]]

Revision as of 21:04, 26 February 2025

Step 1

  • Turn on P2 channel by either double clicking P2, or clicking P2 on keyboard

Step 2

  • Open REAL TIME tab, then select FFR display on. This step will not work if you have not opened P2 yet.

Step 3

  • Once wire is inserted and physician is ready, equalize by double clicking in FFR square.
  • After equalized switch to DFR if not using adenosine.

THIS MAY TAKE A FEW ATTEMPTS FOR PRESSURE TO EQUALIZE, THIS IS NORMAL!!!!!


Step 4

  • Once equalized and on desired measurement selection (FFR or DFR), open MEASUREMENTS tab and select FFR/DFR


Step 5

  • A window will open on the RIGHT sided monitor, all you have to do is click measure, you do NOT have to select a vessel.
  • The measurement should populate under baseline measurements after a few beats.