ECMO: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
No edit summary |
||
Line 35: | Line 35: | ||
*Cor-Knot kit from ECMO cart | *Cor-Knot kit from ECMO cart | ||
*All suture used with Cor-Knot has to be 2-0 or smaller! | *All suture used with Cor-Knot has to be 2-0 or smaller! | ||
==Daniels== | ==Daniels== | ||
===Open=== | |||
*White Nurolon "White Nye" - ECMO cart | |||
*Will secure Prolene purse string with White Nye and carrot top to Arterial Cannula | |||
==Spies== | ==Spies== | ||
*Prep bilateral groins, right radial and right IJ. Patient will need a leave-in swan and RIGHT radial A-line before leaving room. | *Prep bilateral groins, right radial and right IJ. Patient will need a leave-in swan and RIGHT radial A-line before leaving room. | ||
==Antegrade Perfusion== | |||
*Use Arrow short braided sheath on ECMO cart | |||
*Ultrasound and micro puncture used to access SFA below arterial cannula | |||
*Use double-male lure adaptor to attach sheath side-port to arterial cannula vent. | |||
==Bedside Antegrade perfusion== | |||
===Open=== | |||
*ECMO instrument tray | |||
*Pack | |||
*Braided sheath 6 or 7 fr. | |||
*Male to Male adapter | |||
*Micropuncture | |||
*5fr. sheath (Just for wire) | |||
*US probe cover | |||
*0-Ethibond (CT-1) | |||
*Bedside kit | |||
== Spies' Decan without Pre-Closure! == | |||
* 12F Dilator | |||
* 10F Dilator | |||
* 8F Dilator | |||
* 2 Perclose | |||
* Standard J | |||
Clamp arterial cannula, cook needle stick to the cannula. Standard J advanced and cannula removed. 12F dilator inserted, pressure held for 5 or so minutes. Dilator downsized to 10F, pressure held. Perclose times one. 8F dilator advanced and pressure held. Second perclose deployed. |
Revision as of 20:06, 23 January 2025
Note: Apply Foresight to patient before draping to save yourself the hassle later.
Warning: Be prepared to do a Septostomy post ECMO
Equipment
Open from ECMO Cart
- ECMO instruments
- 16/18 Coons dilators - (Only need the 16Fr)
- 20/22 Coons Dilators - (Only need the 20Fr)
- Arterial vented cannula - Normally 16fr.
- Venous cannula - Normally 21fr.
- Irrigation bulb syringe
- ”Carrot tops” Orange vessel tourniquets
- 2-0 Ethibond Multi Pack
Open from room stock
- Micropuncture
- 8fr Pinnacle
- Amplatz SS 7cm tip (x 2 for Spies)
- Cordis 150J
- Probe cover
- (1) Perclose
- (2) 0-prolene (or 2-0, ask)
- (2) Biopatch
Standby
- Amplatz Extra Stiff
- Glidewire
- Alternative sizes of cannulas
- Arrow 5 and 6fr braided sheaths for antegrade bypass
- Male-Male luer adapter for antegrade bypass
Cor-Knot
- Cor-Knot kit from ECMO cart
- All suture used with Cor-Knot has to be 2-0 or smaller!
Daniels
Open
- White Nurolon "White Nye" - ECMO cart
- Will secure Prolene purse string with White Nye and carrot top to Arterial Cannula
Spies
- Prep bilateral groins, right radial and right IJ. Patient will need a leave-in swan and RIGHT radial A-line before leaving room.
Antegrade Perfusion
- Use Arrow short braided sheath on ECMO cart
- Ultrasound and micro puncture used to access SFA below arterial cannula
- Use double-male lure adaptor to attach sheath side-port to arterial cannula vent.
Bedside Antegrade perfusion
Open
- ECMO instrument tray
- Pack
- Braided sheath 6 or 7 fr.
- Male to Male adapter
- Micropuncture
- 5fr. sheath (Just for wire)
- US probe cover
- 0-Ethibond (CT-1)
- Bedside kit
Spies' Decan without Pre-Closure!
- 12F Dilator
- 10F Dilator
- 8F Dilator
- 2 Perclose
- Standard J
Clamp arterial cannula, cook needle stick to the cannula. Standard J advanced and cannula removed. 12F dilator inserted, pressure held for 5 or so minutes. Dilator downsized to 10F, pressure held. Perclose times one. 8F dilator advanced and pressure held. Second perclose deployed.