Created page with "{{Note| Apply Foresight to patient before draping to save yourself the hassle later. }} {{Warn|Be prepared to do a Septostomy post ECMO }} ==Equipment== ===Open from ECMO Cart=== *ECMO instruments *16/18 Coons dilators - (Only need the 16Fr) *20/22 Coons Dilators - (Only need the 20Fr) *Arterial vented cannula - Normally 16fr. *Venous cannula - Normally 21fr. *Irrigation bulb syringe *”Carrot tops” Orange vessel tourniquets *2-0 Ethibond Multi Pack ===Open f..."
 
No edit summary
Line 31: Line 31:
*Arrow 5 and 6fr braided sheaths for antegrade bypass
*Arrow 5 and 6fr braided sheaths for antegrade bypass
*Male-Male luer adapter for antegrade bypass
*Male-Male luer adapter for antegrade bypass
===Cor-Knot===
*Cor-Knot kit from ECMO cart
*All suture used with Cor-Knot has to be 2-0 or smaller!
==Daniels==
==Spies==
*Prep bilateral groins, right radial and right IJ. Patient will need a leave-in swan and RIGHT radial A-line before leaving room.

Revision as of 20:03, 23 January 2025

Note Note: Apply Foresight to patient before draping to save yourself the hassle later.

Note Warning: Be prepared to do a Septostomy post ECMO

Equipment

Open from ECMO Cart

  • ECMO instruments
  • 16/18 Coons dilators - (Only need the 16Fr)
  • 20/22 Coons Dilators - (Only need the 20Fr)
  • Arterial vented cannula - Normally 16fr.
  • Venous cannula - Normally 21fr.
  • Irrigation bulb syringe
  • ”Carrot tops” Orange vessel tourniquets
  • 2-0 Ethibond Multi Pack

Open from room stock

  • Micropuncture
  • 8fr Pinnacle
  • Amplatz SS 7cm tip (x 2 for Spies)
  • Cordis 150J
  • Probe cover
  • (1) Perclose
  • (2) 0-prolene (or 2-0, ask)
  • (2) Biopatch

Standby

  • Amplatz Extra Stiff
  • Glidewire
  • Alternative sizes of cannulas
  • Arrow 5 and 6fr braided sheaths for antegrade bypass
  • Male-Male luer adapter for antegrade bypass

Cor-Knot

  • Cor-Knot kit from ECMO cart
  • All suture used with Cor-Knot has to be 2-0 or smaller!


Daniels

Spies

  • Prep bilateral groins, right radial and right IJ. Patient will need a leave-in swan and RIGHT radial A-line before leaving room.