-A-line, OR 4F in LFA for ACT's | -A-line, OR 4F in LFA for ACT's | ||
|} | |} | ||
===== <nowiki>**</nowiki> Research case, they will bring valve supplies** Supplies do not scan , keep stickers in book! ===== | |||
== '''Timeout Watch Out''' - IS THERE A DEVICE ACROSS THE TRICUSPID VALVE? == | |||
'''Sheaths''' | '''Sheaths''' | ||
* | * Micro puncture | ||
* 8F Sheath for dilator only | * 8F Sheath for dilator only | ||
* 4F Pinnacle (not needed if anesthesia has A-line.) | * 4F Pinnacle (not needed if anesthesia has A-line.) | ||
Line 49: | Line 51: | ||
- Blunt tip forceps & Scissors | - Blunt tip forceps & Scissors | ||
===== Steps ===== | |||
- Patient prepped standard structural way. | |||
[[Category:Procedures]] | [[Category:Procedures]] |
Revision as of 23:20, 11 September 2024
Anesthesia | Imaging | Access |
---|---|---|
Stablizer | 85 cm from zyphoid | |
General | Fluoro,
Tee |
Typical primary access on RFV
-A-line, OR 4F in LFA for ACT's |
** Research case, they will bring valve supplies** Supplies do not scan , keep stickers in book!
Timeout Watch Out - IS THERE A DEVICE ACROSS THE TRICUSPID VALVE?
Sheaths
- Micro puncture
- 8F Sheath for dilator only
- 4F Pinnacle (not needed if anesthesia has A-line.)
Wires
- .032 x 260 "J" wire
- Safarii
Misc. Supplies
- (2) Perclose
- 8.5F Agilis
- Nu-knit
- Probe Cover
- Alligator cables for pacing
- Suture removal kit (for back table)
BACK TABLE SETUP
- Table cover
- 11 Blade
- 4 Bowls
- Normal room temp saline 3.5L
- Heparinized Saline .5L
- Blunt tip forceps & Scissors
Steps
- Patient prepped standard structural way.