(Created page with "work in progress General: -Consent - Electrophysiology Study with Catheter Ablation -Patient intubated/paralyzed (long acting) throughout procedure -No foley -No isuprel -Antibiotics per woods protocol (only w/device, implants etc.) -Fluoro -Goal ACT>300 -McKesson leads, Anesthesia leads, Farapulse leads (slaved into Claris 12 lead by rep) -2 Grounds, Abbott patches, regular defib pad From Pyxis: -protamine if patient is going home. no protamine if patient is...")
 
No edit summary
Line 16: Line 16:
-Goal ACT>300
-Goal ACT>300


-McKesson leads, Anesthesia leads, Farapulse leads (slaved into Claris 12 lead by rep)
-McKesson leads, Anesthesia leads, Farapulse leads (slaved into Claris 12 lead by rep - though started using claris directly 5/2024)


-2 Grounds, Abbott patches, regular defib pad
-2 Grounds, Abbott patches, regular defib pad
Line 55: Line 55:


-Red (pressure bag, no pump)
-Red (pressure bag, no pump)




Charting ablation in Cupid:
Charting ablation in Cupid:
System used - freetext Farapulse
System used - freetext Farapulse
Energy - freetext PFA
Energy - freetext PFA
Number of energy applications - # per Ablation site
Number of energy applications - # per Ablation site




Line 69: Line 74:
8F pinnacle (ICE)
8F pinnacle (ICE)


8f locking sheath (Inquiry CS)
8f locking sheath (duodeca)


11f Pinnacle (or 10.5f fast) ---> Versacross ---> Faradrive ---> Rosen wire loaded into Farawave
11f Pinnacle (or 10.5f fast) ---> Versacross ---> Faradrive ---> Farawave  




Line 83: Line 88:
8f Acunav (non-soundstar)
8f Acunav (non-soundstar)


Inquiry + Cable
Duodeca + Cable


Advisor + Cable
Advisor + Cable
Line 116: Line 121:
vascade x3
vascade x3


Rosen wire
Rosen wire w/3mm J (though md wants to try the pigtailwire from VCkit)


created 5/7/24 zr/ac
created 5/7/24 ac/zr

Revision as of 10:18, 17 June 2024

work in progress General:

-Consent - Electrophysiology Study with Catheter Ablation

-Patient intubated/paralyzed (long acting) throughout procedure

-No foley

-No isuprel

-Antibiotics per woods protocol (only w/device, implants etc.)

-Fluoro

-Goal ACT>300

-McKesson leads, Anesthesia leads, Farapulse leads (slaved into Claris 12 lead by rep - though started using claris directly 5/2024)

-2 Grounds, Abbott patches, regular defib pad


From Pyxis:

-protamine if patient is going home. no protamine if patient is staying. (5/7/24)

-famotidine - less esophageal risk? check before pulling.

-bupivicaine

- (2x) heparin vials (10,000 u/10mL)

- (1x) heparin premixed (25,000/250mL)



Bags:

-500mL NS for ACT line

-1L LR anesthesia

-(1x) 500mL NS with 5000u heparin for the table

-(1x) 500mL NS with 5000u heparin for red line

-(1x prepulled from pyxis) heparin drip for blue line

Lines:

-ACT

-Blue (heparin drip)

-Red (pressure bag, no pump)


Charting ablation in Cupid:

System used - freetext Farapulse

Energy - freetext PFA

Number of energy applications - # per Ablation site



Drop List: Sheaths:

8F pinnacle (ICE)

8f locking sheath (duodeca)

11f Pinnacle (or 10.5f fast) ---> Versacross ---> Faradrive ---> Farawave



Rampart + Radpad

2x yellow filters

8f Acunav (non-soundstar)

Duodeca + Cable

Advisor + Cable

short versacross kit - 180cm pigwire

Light handle cover



1x Ice Sleeve

2x Access Sleeve (tte/linear)

ACT line

2x pressure lines (label blue & red)

Ioban

8F pinnacle

8f locking sheath

11f pinnacle (or 10.5f fastcath)

driver/0pds/stopcock for woggle

chloraprep

vascade x3

Rosen wire w/3mm J (though md wants to try the pigtailwire from VCkit)