TMVI - Encircle Sapien M3: Difference between revisions
From Bay Area Structural Heart Wiki
(Created page with "This will be the Sapien M3 page.") |
No edit summary |
||
Line 1: | Line 1: | ||
DANIELS | |||
Sapien M3 | |||
'''GENERAL ANESTHESIA, TEE, BIPLANE''' | |||
'''ACCESS 2 FEMORAL VEINS (ESHEATH AND PACER), 1 FEMORAL ARTERY''' | |||
'''TRIPLE LUMEN AND RADIAL ARTERY ACCESS BY ANESTHESIA''' | |||
'''STABILIZER PLACEMENT PRIOR TO DRAPING''' | |||
=SHEATHS:= | |||
*(1) 5F MICROPUNCTURE KIT | |||
*(1) 6F 10CM SHEATH | |||
*(1) 8F 10CM SHEATH | |||
*(1) 8F 35CM BRITE TIP SHEATH | |||
*(1) 9F 10CM SHEATH | |||
*(1) 8.5F 63CM SL1 TRANSSEPTAL SHEATH | |||
=WIRES:= | |||
=='''Open'''== | |||
*(2) .035 150CM J WIRE | |||
*(1) .035 260CM J WIRE | |||
*(1) SAFARI 2 WIRE | |||
*(1) AMPLATZ SUPER STIFF 1CM TIP | |||
=='''Standby'''== | |||
*Baylis ProTrack wire | |||
=CATHETERS= | |||
=='''Open'''== | |||
*6F STRAIGHT PIGTAIL 110CM | |||
*6F ANGLED PIGTAIL 110CM | |||
*5F MPA1 | |||
=='''Standby'''== | |||
*6F MPA1 125cm | |||
=PACING= | |||
*5F JOSEPHSON PACING CATHETER | |||
*EP QUAD CABLE | |||
*DUAL RED/BLACK REMINGTON PACING CABLE | |||
*(2) PACER BOXES | |||
=VALVE TABLE= | |||
*(5) STERILE BOWLS (2- 0.9% SALINE, 1-HEP SALINE, 1-15% CONTRAST MIX) | |||
*(1) 90 INCH DRAPE | |||
*1L NORMAL SALINE IRRIGATION | |||
*500ML HEPARINIZED SALINE | |||
*(2) 60ML SYRINGE | |||
*(1) HIGH PRESSURE STOPCOCK | |||
*(1) 20ML SYRINGE | |||
*30ML CONTRAST (MIX WITH 170ML HEP SALINE FOR 15% CONCENTRATION) | |||
*12 INCH EXTENSION | |||
*ROTO TUBING | |||
=='''VALVE- SUPPLIED BY EDWARDS'''== | |||
*INDEFLATOR | |||
*CRIMPER | |||
*E SHEATH | |||
*DELIVERY SYSTEM | |||
*VALVE | |||
=TRANS SEPTAL= | |||
*BAYLIS NEEDLE LARGE CURVE C1 | |||
*BAYLIS MACHINE | |||
*BAYLIS CABLE | |||
*GROUNDING PAD | |||
=SEPTAL DILATION= | |||
=='''Open'''== | |||
*14X4 ATLAS GOLD- 135 CM SHAFT PREFERRED | |||
*PRESTO INDEFLATOR | |||
=='''Standby'''== | |||
*TORQUEVUE 45DEGREE DELIVERY SYSTEM (9F OR 10F) | |||
*10MM, 14MM, 18MM SEPTAL OCCLUDER | |||
=ADDITIONAL SUPPLIES= | |||
=='''Open'''== | |||
*(2) PERCLOSE | |||
*3 PORT MANIFOLD | |||
*ACIST- 70/30 MIX | |||
*PROBE COVER | |||
*DUMP BAG | |||
*25g 1½ INCH NEEDLE | |||
*STAT LOCK WITH STRING SECUREMENT | |||
*PROLENE 2-0 | |||
*TAVR INSTRUMENT TRAY | |||
*CHLORAPREP AND PREP TRAY | |||
*DERMABOND | |||
*(2) FLUORO DRAPES (FOR LATERAL BASE AND ANESTHESIA POLE) | |||
*36X36 BANDBAG (FOR LATERAL DETECTOR) | |||
*1L HEP SALINE ON PRESSURE BAG HUNG FROM IVY POLE | |||
*That's right, I said IVY pole. | |||
=MEDICATIONS= | |||
*BUPIVACAINE/LIDOCAINE MIX | |||
*EPINEPHRINE 10MCG/ML IN RED SYRINGE ON BACK TABLE | |||
*ROTAGLIDE 10ML (PRE VALVE) | |||
=STANDBY EMERGENCY ECMO SUPPLIES= | |||
*16F ARTERIAL CANNULA | |||
*21F VENOUS CANNULA | |||
*(2) PERCLOSE | |||
*ECMO INSTRUMENT TRAY | |||
*AMPLATZ SUPER STIFF | |||
=STANDBY PERICARDIOCENTESIS SUPPLIES= | |||
*EYE DRAPE | |||
*CHIBA NEEDLE 10CM 18G | |||
*SKATER INTRODUCER SET | |||
*BARD LOCKING PIGTAIL 6F OR 8F | |||
*(3) 60ML SYRINGES | |||
*STOPCOCK | |||
*AMPLATZ SUPER STIFF 180CM | |||
=STANDBY PARAVALVULAR LEAK REPAIR SUPPLIES= | |||
'''SEE MITRAL PARAVALVULAR LEAK REPAIR''' | |||
*AMPLATZER VASCULAR PLUG 2 SIZE 8-18 AVAILABLE | |||
=PROCEDURE:= | |||
*Patient prepped from '''navel to knees''' in sterile fashion. | |||
*ISOCENTER XRAY | |||
*Access obtained with u/s | |||
Typical sites: | |||
***6F LFA for Pigtail | |||
***8F 35cm LFV for transcutaneous pacing | |||
***20F RFV E-sheath for valve delivery | |||
****Dilate with 8F | |||
****Preclose | |||
*Pigtail advanced through 6F sheath and LVgram performed | |||
*Mitral annulus denoted on live monitor screen in A plane | |||
*SL1 inserted over .032 kit wire | |||
*Baylis needle inserted through SL1 and atrial septum puncture performed. | |||
*Needle removed | |||
*5F MPA and standard J wire advanced to pulmonary vein | |||
*1cm Amplatz Super Stiff advanced and MPA and SL1 removed | |||
*Edwards Guide Sheath (GS) advanced | |||
*Dock Delivery System (DDS) advanced through GS | |||
*Dock deployed in a timely fashion ;) | |||
*DDS removed and MPA inserted over exchange J for valve crossing | |||
*J wire exchanged for Safari 2 and MPA removed | |||
*14mm balloon advanced for septostomy | |||
*Balloon and GS removed | |||
*Valve delivery system advanced | |||
*When valve is in place, temp pacer inserted and tested. | |||
*Valve deployed. Septum evaluated with echo for shunting and PVL. | |||
*Delivery system removed and MPA inserted for LA pressure measurement. | |||
*All equipment removed | |||
*Preclosure closed and stitch to RFV | |||
*Perclose LFA | |||
*8F vein exchanged for 9F 10cm sheath | |||
*Secure sheath with Statlock. Tegaderm both sides. | |||
<br /> |
Revision as of 15:02, 25 September 2019
DANIELS
Sapien M3
GENERAL ANESTHESIA, TEE, BIPLANE
ACCESS 2 FEMORAL VEINS (ESHEATH AND PACER), 1 FEMORAL ARTERY
TRIPLE LUMEN AND RADIAL ARTERY ACCESS BY ANESTHESIA
STABILIZER PLACEMENT PRIOR TO DRAPING
SHEATHS:
- (1) 5F MICROPUNCTURE KIT
- (1) 6F 10CM SHEATH
- (1) 8F 10CM SHEATH
- (1) 8F 35CM BRITE TIP SHEATH
- (1) 9F 10CM SHEATH
- (1) 8.5F 63CM SL1 TRANSSEPTAL SHEATH
WIRES:
Open
- (2) .035 150CM J WIRE
- (1) .035 260CM J WIRE
- (1) SAFARI 2 WIRE
- (1) AMPLATZ SUPER STIFF 1CM TIP
Standby
- Baylis ProTrack wire
CATHETERS
Open
- 6F STRAIGHT PIGTAIL 110CM
- 6F ANGLED PIGTAIL 110CM
- 5F MPA1
Standby
- 6F MPA1 125cm
PACING
- 5F JOSEPHSON PACING CATHETER
- EP QUAD CABLE
- DUAL RED/BLACK REMINGTON PACING CABLE
- (2) PACER BOXES
VALVE TABLE
- (5) STERILE BOWLS (2- 0.9% SALINE, 1-HEP SALINE, 1-15% CONTRAST MIX)
- (1) 90 INCH DRAPE
- 1L NORMAL SALINE IRRIGATION
- 500ML HEPARINIZED SALINE
- (2) 60ML SYRINGE
- (1) HIGH PRESSURE STOPCOCK
- (1) 20ML SYRINGE
- 30ML CONTRAST (MIX WITH 170ML HEP SALINE FOR 15% CONCENTRATION)
- 12 INCH EXTENSION
- ROTO TUBING
VALVE- SUPPLIED BY EDWARDS
- INDEFLATOR
- CRIMPER
- E SHEATH
- DELIVERY SYSTEM
- VALVE
TRANS SEPTAL
- BAYLIS NEEDLE LARGE CURVE C1
- BAYLIS MACHINE
- BAYLIS CABLE
- GROUNDING PAD
SEPTAL DILATION
Open
- 14X4 ATLAS GOLD- 135 CM SHAFT PREFERRED
- PRESTO INDEFLATOR
Standby
- TORQUEVUE 45DEGREE DELIVERY SYSTEM (9F OR 10F)
- 10MM, 14MM, 18MM SEPTAL OCCLUDER
ADDITIONAL SUPPLIES
Open
- (2) PERCLOSE
- 3 PORT MANIFOLD
- ACIST- 70/30 MIX
- PROBE COVER
- DUMP BAG
- 25g 1½ INCH NEEDLE
- STAT LOCK WITH STRING SECUREMENT
- PROLENE 2-0
- TAVR INSTRUMENT TRAY
- CHLORAPREP AND PREP TRAY
- DERMABOND
- (2) FLUORO DRAPES (FOR LATERAL BASE AND ANESTHESIA POLE)
- 36X36 BANDBAG (FOR LATERAL DETECTOR)
- 1L HEP SALINE ON PRESSURE BAG HUNG FROM IVY POLE
- That's right, I said IVY pole.
MEDICATIONS
- BUPIVACAINE/LIDOCAINE MIX
- EPINEPHRINE 10MCG/ML IN RED SYRINGE ON BACK TABLE
- ROTAGLIDE 10ML (PRE VALVE)
STANDBY EMERGENCY ECMO SUPPLIES
- 16F ARTERIAL CANNULA
- 21F VENOUS CANNULA
- (2) PERCLOSE
- ECMO INSTRUMENT TRAY
- AMPLATZ SUPER STIFF
STANDBY PERICARDIOCENTESIS SUPPLIES
- EYE DRAPE
- CHIBA NEEDLE 10CM 18G
- SKATER INTRODUCER SET
- BARD LOCKING PIGTAIL 6F OR 8F
- (3) 60ML SYRINGES
- STOPCOCK
- AMPLATZ SUPER STIFF 180CM
STANDBY PARAVALVULAR LEAK REPAIR SUPPLIES
SEE MITRAL PARAVALVULAR LEAK REPAIR
- AMPLATZER VASCULAR PLUG 2 SIZE 8-18 AVAILABLE
PROCEDURE:
- Patient prepped from navel to knees in sterile fashion.
- ISOCENTER XRAY
- Access obtained with u/s
Typical sites:
- 6F LFA for Pigtail
- 8F 35cm LFV for transcutaneous pacing
- 20F RFV E-sheath for valve delivery
- Dilate with 8F
- Preclose
- Pigtail advanced through 6F sheath and LVgram performed
- Mitral annulus denoted on live monitor screen in A plane
- SL1 inserted over .032 kit wire
- Baylis needle inserted through SL1 and atrial septum puncture performed.
- Needle removed
- 5F MPA and standard J wire advanced to pulmonary vein
- 1cm Amplatz Super Stiff advanced and MPA and SL1 removed
- Edwards Guide Sheath (GS) advanced
- Dock Delivery System (DDS) advanced through GS
- Dock deployed in a timely fashion ;)
- DDS removed and MPA inserted over exchange J for valve crossing
- J wire exchanged for Safari 2 and MPA removed
- 14mm balloon advanced for septostomy
- Balloon and GS removed
- Valve delivery system advanced
- When valve is in place, temp pacer inserted and tested.
- Valve deployed. Septum evaluated with echo for shunting and PVL.
- Delivery system removed and MPA inserted for LA pressure measurement.
- All equipment removed
- Preclosure closed and stitch to RFV
- Perclose LFA
- 8F vein exchanged for 9F 10cm sheath
- Secure sheath with Statlock. Tegaderm both sides.