Mitral Clip: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
No edit summary
Line 1: Line 1:
== DRAFT ==
==DRAFT==
{| class="wikitable"
{| class="wikitable"
! colspan="6" |Mitral Clip
! colspan="6" |Mitral Clip
Line 13: Line 13:
*Femoral Vein x2
*Femoral Vein x2
|
|
*Aspirin
|}
|}


Line 68: Line 67:
====Open====
====Open====


*ITEMS TO ALWAYS OPEN
*(1) Cath pack
*(2) Items in ALL CAPS
*(1) Lift / Footstool / Support base
*(quantity) ITEM
*(1) Clear Plexiglass base support
*(1) Sterile System Stabilizer
*(1) Baylis generator
*(1)  


====Standby====
====Standby====
Line 87: Line 89:
  Start line with a space
  Start line with a space
       More spaces inset more
       More spaces inset more
 
   
  [[Category:Procedures]]
    [[Category:Procedures]]


APPROVED: MD initials MM/YY
APPROVED: MD initials MM/YY

Revision as of 15:12, 4 October 2019

DRAFT

Mitral Clip
Anesthesia Imaging Access Pre-Procedure
General Fluoro, TEE, and ICE
  • Femoral Vein x2


Mitral Clip

Open:

(1)   Liter pressure bags

(2)   108cm High pressure Ivy Lines

(5) High Pressure Stopcock

(1) 8F Dilator

(1) 20F Dilator for CS

(1) 5F MPA2

(1) Amplatz SuperStiff 1cm Tip

(1) .035 X 260 “J” wire

(1) .035 x 150 “J” wire

(1) 8.5 F SL1

(1) Baylis large curve transeptal needle and cable

(1) Perclose

(1) 2-0 Prolene suture

(1) Probe cover

(2) 60cc syringe

(1) 12” extension (Keep stickers to label flush tubing)

(3) Sterile towel packs for back table

(1) 3 port manifold to measure LA pressure only

(1) Fluoro cover to anesthesia Ivy pole or moving xray shield

Procedure:

Prior to prepping patient, place MitralClip Plate (plexiglass board – hinge side up) under patient’s leg off center to the right. Place MitralClip Lift (step stool) over patients right leg, angled down towards the head. Measure 80cm from mid-sternum to leading edge of stool. Make sure there is direct contact between stool and plate.

Prep patient and drape per usual. Make sure to drape anesthesia pole and or xray shield.

Main Header e.g. wires,sheaths:

Open

  • (1) Cath pack
  • (1) Lift / Footstool / Support base
  • (1) Clear Plexiglass base support
  • (1) Sterile System Stabilizer
  • (1) Baylis generator
  • (1)

Standby

  • STANDBY ITEMS
  • GO HERE


Procedure

  • Bullet list of typical procedure steps
  • More steps
Inset alternative steps
or notes
Start line with a space
     More spaces inset more

APPROVED: MD initials MM/YY