Mitral Clip: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
No edit summary |
||
Line 17: | Line 17: | ||
===Procedural Gems:=== | ===Procedural Gems:=== | ||
*When using the PDOII plug In MitraClip setting use TorqView 180 smaller cable | *When using the PDOII plug In MitraClip setting use TorqView 180 (smaller cable needed) | ||
===Big Equipment:=== | ===Big Equipment:=== | ||
Line 71: | Line 71: | ||
*Perclose | *Perclose | ||
*0 Prolene suture on a CT-1 needle (Spies) | *0 Prolene suture on a CT-1 needle (Spies) | ||
*60cc syringe | *(2) 60cc syringe | ||
*12” extension (Keep stickers to label flush tubing) | *12” extension (Keep stickers to label flush tubing) | ||
*(2) Sterile towel packs for back table | *(2) Sterile towel packs for back table | ||
Line 89: | Line 89: | ||
===Non-Sterile Supplies=== | ===Non-Sterile Supplies=== | ||
*( | *(2) Liter bags of heparinized saline | ||
*( | *(2) Liter sized pressure bags | ||
==Procedure== | ==Procedure== | ||
Line 121: | Line 121: | ||
BACK TABLE SETUP | BACK TABLE SETUP | ||
*108" Tubing spikes handed off to circulator, keep rollers on sterile field. | * 108" Tubing spikes handed off to circulator, keep rollers on sterile field. | ||
*Follow Rep instructions for device prep. | *Follow Rep instructions for device prep. | ||
Revision as of 00:49, 6 November 2019
DRAFT
Mitral Clip | |||||
---|---|---|---|---|---|
Anesthesia | Imaging | Access | Pre-Procedure | ||
General | Fluoro, TEE, and ICE |
Femoral Vein |
Procedural Gems:
- When using the PDOII plug In MitraClip setting use TorqView 180 (smaller cable needed)
Big Equipment:
Open
- Cath pack
- Lift / Footstool / Support base
- Clear Plexiglass base support
- Sterile System Stabilizer
- Baylis generator (Mitral only)
- ICE 8fr. and SwiftLink cover (Tricuspid only)
- 9fr x 23cm Pinnacle for ICE
Baylis/transseptal:
Not used for Tricuspid clipping
Open
- SL1 TransSeptal sheath
- Large curve Baylis needle
- Baylis cable
- Grounding pad
Standby
- BRK1 transseptal needle (Rarely used)
Wires:
Open
- Cordis 260J
- Cordis 150J
- Amplatz SuperStiff 1cm Tip
Standby
- .032 X 260 “J” wire
- .014 X 190 BMW wire
Catheters:
Open
- 5F MPA2
Misc supplies:
Open
- (2) 108cm High pressure IV Lines
- (5) High Pressure Stopcock
- 8F Sheath for dilator - (also used as a backup sheath)
- Perclose
- 0 Prolene suture on a CT-1 needle (Spies)
- (2) 60cc syringe
- 12” extension (Keep stickers to label flush tubing)
- (2) Sterile towel packs for back table
- 3 port manifold to measure LA pressure only (not needed for Tricuspid)
- Fluoro cover to anesthesia IV pole or moving xray shield
- 20F Dilator
- Needle driver
- Micropuncture
- U/S probe cover
- dump bag
Standby
- Additional Perclose
- Pericardiocentesis supplies
Non-Sterile Supplies
- (2) Liter bags of heparinized saline
- (2) Liter sized pressure bags
Procedure
- Anesthesia will obtain radial art line prior to draping.
- Pre procedure TEE imaging performed
- Prior to prepping patient, place MitraClip Plate (plexiglass board – hinge side up) under patient’s leg off center to the right. Place MitralClip Lift (step stool) over patients right leg, angled down towards the head. Measure 80cm from mid-sternum to leading edge of stool. Make sure there is direct contact between stool and plate.
- Prep patient and drape per usual. Drape anesthesia pole and/or xray shield.
- Obtain RFV access, dilate with 8F, preclose. Hand MD .032wire loaded into micropuncture sheath after Perclose deployment.
- SL1 introduced over .032 wire.
- Baylis needle inserted and echo imaging performed. Transseptal puncture performed.
- Baylis needle removed. .032wire inserted. Wire and dilator removed.
- MPA and Jwire inserted to pulmonary vein. LA pressure measured. Give MD Amplatz SuperStiff to shape.
- Jwire removed and Amplatz SuperStiff advanced.
- MPA removed.
- SL1 removed
- 20F dilator inserted and removed
- MitraClip Guide inserted.
- Setup stabilizer and silicone sheet. Hand MD 60ml syringe with short extension tubing for aspiration.
- Amplatz wire and dilator removed. Clip inserted.
- Drips should be maintained at 1 drip every 4 seconds. If bubbles are seen on echo, turn drips down.
- When clip is being deployed turn room lights up and drips up. Place bowl below to catch drips while deploying.
- Hand MD 60ml syringe with extension tubing for aspiration while delivery system is removed.
- LA pressure measured. Set up for subsequent clips if needed.
- Prolene used for groin closure.
BACK TABLE SETUP
- 108" Tubing spikes handed off to circulator, keep rollers on sterile field.
- Follow Rep instructions for device prep.
APPROVED: MD initials MM/YY