Impella: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
|||
Line 84: | Line 84: | ||
*IABP secured with suture and site dressed. | *IABP secured with suture and site dressed. | ||
*Radial sheath removed and TR band applied. | *Radial sheath removed and TR band applied. | ||
[[Category:Procedures]] |
Revision as of 20:47, 11 October 2019
Daniels single access Impella - DRAFT
Equipment
- 16F Dryseal
- 8fr guide
- 6fr Shuttle 90cm
Hybrid OR Impella via Subclavian conduit
Equipment
- Pack*
- JR4
- Angled Pigtail
- .025 exchange for BP removal
- Perclose
- .035 exchange
- 6-8.5 statlock
- 18” cover for foot pedal*
- 500ml bag heparinized saline
- *Trust me on this...
Procedure
- Enter an order for CCL in EPIC
- Open x-ray and McKesson using entered order.
- Procedure in McKesson is “(33990)INSERT VAD ARTERY ACCESS”
- Free-Text note that McKesson is open for billing and equipment only, not procedural charting. See Anesthesia and OR nurses notes.
- Put input 4in1 B on V-integration 2 for TTE on the fluoro screen.
- Open and cut femoral drape across entire width at the level of the femoral fenestrations.
- Place over legs after surgical draping using remaining adhesive area to secure.
- Put OR light camera in place to see the surgical site to monitor progress from control room.
- Tuck surgical drapes under mattress to expose controls.
- Ask OR for 1 pack Ray-Tec sponges for wipes.
- Do the thing (Insert 5.0 Impella)
- Remove the other thing (ECMO / Balloon pump / Impella)
- Statlock is for the chest/abdomen loop of the Impella similar to Leave in swan.
- Document times and doses in McKesson - Write down for book.
- Save 2 stickers
- Close x-ray and archive to EMIM_CATH
- Capture charges for McKesson
- Generate and sign McKesson, then print device list in CCL.
- Sticker goes in room 6 book for tracking, second sticker is for device list
- Bill device list in Epic
Axillary Impella/IABP - DRAFT
If left subclavian is occluded or otherwise inaccessible procedure is aborted. IABP should not be inserted through right subclavian per CT surgery.
Equipment
- Micropuncture
- (2) 4/5f slender radial sheath
- Amplatz extra stiff 260cm wire
- .035 150cm Jwire
- 5f pigtail catheter
- 5f JR4 diagnostic catheter
- IABP kit
- 8f 23cm safesheath (ep)(Romick) / 8fr x 23 brite-tip sheath (Spies)
- Perclose
- TR band
- Biopatch
- 0-Ethibond suture
Procedure
- Prep left side neck and chest (ear to nipple and shoulder to midline).
- Note: The access site is not an absolute (may become axillary or subclavicular) hence the large site prep.
- Prep left radial.
- Left radial accessed and 4/5f slender sheath inserted.
- Pigtail advanced and angiogram performed. Pigtail left in place and used as reference for subclavian access.
- Subclavian accessed with ultrasound and fluoro guidance using micropuncture kit and radial sheath inserted.
- Pigtail catheter removed.
- J wire and JR4 inserted and directed to descending aorta.
- Catheter and sheath removed, 8f dilator inserted, and perclose advanced for pre-closure.
- Amplatz wire advanced, perclose removed, 8f sheath advanced.
- IABP inserted.
- Sheath peeled away. (Romick)
- IABP secured with suture and site dressed.
- Radial sheath removed and TR band applied.