Cryo Setup: Difference between revisions

From Bay Area Structural Heart Wiki
Rosenbzj (talk | contribs)
No edit summary
Rosenbzj (talk | contribs)
No edit summary
 
Line 5: Line 5:
* Electrical Umbilical (Sterile core)
* Electrical Umbilical (Sterile core)
* Coaxial Umbilical  (Sterile core)
* Coaxial Umbilical  (Sterile core)
* 8F sheath, usually a long sheath like an SR0


'''<u>Steps</u>'''
'''<u>Steps</u>'''

Latest revision as of 02:15, 8 October 2025

Cryo Equipment/Supplies

  • Cryo Console (Located in sterile core)
  • Cryo Catheter (4,6,or 8mm)
  • Electrical Umbilical (Sterile core)
  • Coaxial Umbilical (Sterile core)

Steps

  • Get cryo console and position towards the foot of table on scrub side
  • Connect Scavenging Hose (white hose on console) to vacuum port on boom near anesthesia
  • Plug in power cable
  • Open refrigerant tank in console (Lefty Loosey)
  • Power up console (It is recommended that power is on at least 15mins prior to starting procedure)
  • Check for error messages (If yes, check all connections and refrigerant tank is open and full. Reboot)
  • After power up - Complete patient info screen
  • Connect Auto-connection box (located inside the console) to the front of the console
  • Plug pin connector box (located in room at bottom of 3rd cabinet) into 20pole B
  • Connect Auto connection box pins in 20pole B 1 through 4 (Distal to Proximal)
  • Connect Electrical Umbilical to Auto connection box (From sterile table to ACB)
  • Connect Coaxial Umbilical to front of Console (From Sterile table to Console)
  • Connect Catheter to Electrical and coaxial umbilicals
  • Ensure appropriate signal routing on Prucka


Misc error codes:

-Message 50005

1) Ablation generator has not been powered off

2) Distal pin has not been disconnected prior to freeze

3) Blood in Cryo catheter