| Line 37: | Line 37: | ||
* Wholey / Versacore | * Wholey / Versacore | ||
=== Amulet supplies: === | === Amulet supplies: (supplies provided by Rep.) === | ||
* 12 Fr or 14 Fr Tourque View Delivery (Open as requested after TEE) | * 12 Fr or 14 Fr Tourque View Delivery (Open as requested after TEE) | ||
| Line 51: | Line 51: | ||
# Baby J wire from sheath | # Baby J wire from sheath | ||
# 8 Fr dilator | # 8 Fr dilator | ||
# 6 Fr Perclose pre-close RFV | # 6 Fr Perclose pre-close RFV. | ||
# .035 J wire from versacross system advanced thru perclose for Baylis sheath advancement. | # .035 J wire from versacross system advanced thru perclose for Baylis sheath advancement. | ||
# .035 J wire removed, VersaCross RF pigtail wire advanced for transeptal puncture. | # .035 J wire removed, VersaCross RF pigtail wire advanced for transeptal puncture. | ||
| Line 58: | Line 58: | ||
# Transeptal puncture, ACT drawn from LA. | # Transeptal puncture, ACT drawn from LA. | ||
# Measure LA pressure | # Measure LA pressure | ||
# Tourque View Delivery advanced to the LA | # ? Tourque View Delivery advanced to the LA | ||
# Backloaded 5 Fr Str Pigtail advanced to the LAA | # ? Backloaded 5 Fr Str Pigtail advanced to the LAA | ||
# Hook pigtail up to manifold for angiogram of LAA | # ? Hook pigtail up to manifold for angiogram of LAA | ||
# Measure and select size of device | # Measure and select size of device | ||
# Open and prep the device | # Open and prep the device | ||
Revision as of 16:51, 29 May 2025
| Anesthesia | Imaging | Access | Acist/Manifold |
|---|---|---|---|
| Gernerl | Fluoro and TEE (ICE?) | 12 Fr or 14Fr RFV | Manifold with contrast |
| Rep: Monica Tseng 415-816-2877 | |||
Open
- Micropuncture
- 8fr sheath (dilator used pre-Perclose)
- Probe cover
- 1 Perclose
- 0-Prolene or 0-PDS (Spies will add a suture after perclose if site is oozing. If no ooze Exofin.)
- Cordis 150J
- 5fr Straight Pigtail
- 3-port manifold
- Baylis VersaCross Access System (purple box in structural cart)
- Grounding pad
- (Spies - Lido with EPI)
Standby
- 4 or 5fr Pinnacle for Arterial line
- 190 Supracore
- Wholey / Versacore
Amulet supplies: (supplies provided by Rep.)
- 12 Fr or 14 Fr Tourque View Delivery (Open as requested after TEE)
- Amulet device (Size as requested after TEE)
PROCEDURE
- Basic table set up ONLY until it is determined by TEE that case will proceed.
- TEE performed. Procedure aborted if clot
- Prep bilateral groins and drape once probe has been inserted.
- Micropuncture ultrasound-guided RFV access
- 5 Fr Micropuncture sheath
- Baby J wire from sheath
- 8 Fr dilator
- 6 Fr Perclose pre-close RFV.
- .035 J wire from versacross system advanced thru perclose for Baylis sheath advancement.
- .035 J wire removed, VersaCross RF pigtail wire advanced for transeptal puncture.
- TEE & fluoro guided transeptal access & images of Left Atrial Appendage
- Anesthesia gives Heparin IV. ACT GOAL >250-300
- Transeptal puncture, ACT drawn from LA.
- Measure LA pressure
- ? Tourque View Delivery advanced to the LA
- ? Backloaded 5 Fr Str Pigtail advanced to the LAA
- ? Hook pigtail up to manifold for angiogram of LAA
- Measure and select size of device
- Open and prep the device
- Pigtail removed just prior to delivery system insertion.
- Placement confirmed by LA angiogram
- Delivery system removed as MD tamps perclose knot.
- If access is clean, Dermabond. If bleeding, woggle