Septostomy: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
No edit summary
Line 8: Line 8:
* Indiflator
* Indiflator
* Naked 3 port manifold for LA pressure
* Naked 3 port manifold for LA pressure
* Sheath , one French size larger then minimum balloon size
* Balloon, size determined via ECHO


===* NEW - Baylis Transseptal===
===* NEW - Baylis Transseptal===

Revision as of 03:28, 9 March 2024

Z-Med Inflation Chart
Other Balloons Inflation Chart

An atrial Septostomy is performed in the setting of ECMO when there is a concern that Left Atrial pressures are high due to the ECMO flow. Generally done in the setting of ECMO. It can be a stand alone procedure if not performed when ECMO is initiated.

Equipment: Daniels

  • Indiflator
  • Naked 3 port manifold for LA pressure
  • Sheath , one French size larger then minimum balloon size
  • Balloon, size determined via ECHO

* NEW - Baylis Transseptal

  • VersaCross system
  • Grounding pad

STANDBY!

  • SL1 TransSeptal sheath
  • Large curve Baylis needle
  • Baylis cable
  • .032 X 260 “J” wire
  • Amplatz SuperStiff 1cm Tip 260cm
  • 5F MPA2

Equipment: Spies

  • Indiflator
  • Naked 3 port manifold for LA pressure
  • Amplatz SuperStiff 1cm Tip 260cm
  • 5F MPA2

* NEW - Baylis Transseptal

  • VersaCross system
  • Grounding pad

STANDBY!

  • SL1 TransSeptal sheath
  • Large curve Baylis needle
  • Baylis cable
  • .032 X 260 “J” wire

Procedure requires Baylis

  • Baylis generator
  • Grounding Pad

Balloon size determined by MD.

  • Sheath size determined by balloon. Balloon volume listed on Structural Heart Cart, and chart above.
  • Have a 22 Tyshak nucleus on standby for Spies and Daniels



Approved: MD initials MM/YY