Alaris Pump Pressure: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
No edit summary
Line 10: Line 10:
#Select the “Selectable” option
#Select the “Selectable” option
#Using the up and buttons, change the pressure limit to 400 mmHg
#Using the up and buttons, change the pressure limit to 400 mmHg
'''Plum360 Pump Set Up'''
# Select Procedural from drug library
# Select pump A
# Text "IV " and choose IV Fluid Bolus
# Change the Container Volume and Dose to 900mL
# Change rate to 200mL
# Push Start and Click "Yes" to override
# Select Settings and select pressure (Max 15 psi - this part is going to be trial by error)
# Select previous screen and start infusion


[[Category:Setups]]
[[Category:Setups]]

Revision as of 16:49, 9 March 2022

This page is a guide to overriding the pressure settings on the Alaris IV pump. Primarily used for the Sapien M3 setup.

Steps

  1. Press the white “Select Channel” button on the pump module.
  2. Select “Basic Infusion” on the brain.
  3. Change the rate to 200 ml/hr
  4. Change the VTBI to 900ml
  5. Press the options button on the brain at the bottom left of the display.
  6. Select “Pressure Limit”
  7. Select the “Selectable” option
  8. Using the up and buttons, change the pressure limit to 400 mmHg


Plum360 Pump Set Up

  1. Select Procedural from drug library
  2. Select pump A
  3. Text "IV " and choose IV Fluid Bolus
  4. Change the Container Volume and Dose to 900mL
  5. Change rate to 200mL
  6. Push Start and Click "Yes" to override
  7. Select Settings and select pressure (Max 15 psi - this part is going to be trial by error)
  8. Select previous screen and start infusion