Triluminate RHC: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
No edit summary |
||
Line 1: | Line 1: | ||
Equipment: 6F Balloon wedge ; 5/6 Slender | |||
Synopsis of requirements for RHC performed for Triluminate Trial | Synopsis of requirements for RHC performed for Triluminate Trial | ||
Latest revision as of 18:13, 13 August 2021
Equipment: 6F Balloon wedge ; 5/6 Slender
Synopsis of requirements for RHC performed for Triluminate Trial
- PAP measured at end expiration
- 2 samples with less than 10% difference between the two
- PCWP measured at end expiration
- 2 samples, should not differ by more than 3mmHg
- RAP
- 2 samples with less than 10% difference between the two
- Fick method of CO preferred
- Review all tracings- computer generated values should match actual tracings.
Goals for Trial
PAP <70mmHg
PCWP <25mmHg