TAVI-JenaValve: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
No edit summary
Line 6: Line 6:
'''As of 12/11 we will now pace for ALL Jenavalve implants.'''
'''As of 12/11 we will now pace for ALL Jenavalve implants.'''


'''Anesthesia- GA is standard. May consider MAC based on patient.  If MAC, may want ICE.'''
'''Anesthesia- MAC is our new standard. Have ICU on standby.'''  


===Sheaths:===
===Sheaths:===
Line 12: Line 12:
*5Fr Micropunture Kit
*5Fr Micropunture Kit
*6Fr x 10cm sheath
*6Fr x 10cm sheath
*8Fr x 10 cm sheath (Dilator used only)
*8Fr x 10 cm sheath  
*8Fr x 35 cm Brite Tip Sheath
*18F Cook Sheath- on standby
*18F Cook Sheath- on standby


Line 38: Line 37:
*3-port manifold kit
*3-port manifold kit
*Acist hand control
*Acist hand control
*3L cold saline
*1 L warm saline (on standby)
*1L warm saline
*3 L cold saline
*(2) 60ml syringes
*(2) 60ml syringes (for back table)
*(2) 20ml syringes
*(2) 20ml syringes (one for back table one for front)
*stopcock
*stopcock (for back table)
*co-pilot
*co-pilot (for back table)
*108cm tubing
*108cm tubing (for back table)
*500ml hep saline on pressure bag
*500ml hep saline on pressure bag for delivery system
*(3) Perclose
*(3) Perclose
*(5) bowls
*(5) bowls (cold saline in all 5 * NO need for contrast mix on back table *)
*Alligator pacer cable (Unipolar pacing)
*Suture removal kit (for back table)


===Transvenous pacemaker supplies===
===On Standby : Transvenous pacemaker supplies===


*8F Brite tip Sheath
*5Fr Josephson Pacing catheter
*5Fr Josephson Pacing catheter
*Dual Red/Black Pacing Remington cables
*Dual Red/Black Pacing Remington cables

Revision as of 16:46, 10 March 2022


Daniels/Spies - Draft

ACIST A1000V SYRINGE USED WITH 50/50 CONTRAST MIX

As of 12/11 we will now pace for ALL Jenavalve implants.

Anesthesia- MAC is our new standard. Have ICU on standby.

Sheaths:

  • 5Fr Micropunture Kit
  • 6Fr x 10cm sheath
  • 8Fr x 10 cm sheath
  • 18F Cook Sheath- on standby

Wires:

  • (2) Cordis 150J
  • Cordis 260J
  • Safari 2

Catheters:

  • 5Fr MPA
  • 5Fr Angled Pigtail
  • 5F MBP on standby

Medications on table:

  • Lido/Bupivicaine 50/50
  • Epinephrine 10mcg/ml in red medallion syringe (From anesthesia)

Miscellaneous Supplies

  • 3-port manifold kit
  • Acist hand control
  • 1 L warm saline (on standby)
  • 3 L cold saline
  • (2) 60ml syringes (for back table)
  • (2) 20ml syringes (one for back table one for front)
  • stopcock (for back table)
  • co-pilot (for back table)
  • 108cm tubing (for back table)
  • 500ml hep saline on pressure bag for delivery system
  • (3) Perclose
  • (5) bowls (cold saline in all 5 * NO need for contrast mix on back table *)
  • Alligator pacer cable (Unipolar pacing)
  • Suture removal kit (for back table)

On Standby : Transvenous pacemaker supplies

  • 8F Brite tip Sheath
  • 5Fr Josephson Pacing catheter
  • Dual Red/Black Pacing Remington cables
  • EP Quad Cable

PROCEDURAL INFO:

  • Standard TAVR access.
  • Pigtail in contralateral access, injection for coplanar angle.
  • Change injector to 10/10 after coplanar angle established.
  • Pigtail exchanged for MPA
  • Pigtail advanced on primary side for valve crossing over J wire
  • Jwire exchanged for Safari and pigtail removed
  • JV Guide sheath inserted (when ACT above 250)
  • Delivery system attached to pressure bag via co-pilot on front table.
  • Valve inserted and deployed
  • Delivery system pulled back into JVsheath
  • Possible exchange for 18F if concern about pulling safari through/BAV needed