===Equipment:=== | |||
* 260cm J Wire | |||
* 20F Dryseal | |||
* Agilis | |||
* 5F 110cm Shuttle Sheath (dependent on plug size) | |||
* 5F 10cm Pinnacle Sheath | |||
* Straight glidewire (standard) | |||
* 180* torqvue (French size not important here, only using for the cable.)<br /> | |||
=== Procedure for placing a plug between 2 Mitraclip in the same setting: === | |||
* 8F sheath and 260cm J inserted through guide delivery system. | |||
* Guide delivery system removed, 20F dryseal inserted | |||
* Agilis inserted | |||
* Preload Shuttle sheath with glidewire and advance through agilis | |||
* Glidewire used to cross between plugs and removed with shuttle dilator. | |||
* 5F sheath inserted into shuttle in lieu of using torqvue delivery sideport | |||
** '''Transition between shuttle and torqvue insertion isn't flush so occluder starts to deploy if sheath not used''' | |||
<br /> | <br /> |
Revision as of 22:29, 30 June 2020
ADOII
Equipment:
- 260cm J Wire
- 20F Dryseal
- Agilis
- 5F 110cm Shuttle Sheath (dependent on plug size)
- 5F 10cm Pinnacle Sheath
- Straight glidewire (standard)
- 180* torqvue (French size not important here, only using for the cable.)
Procedure for placing a plug between 2 Mitraclip in the same setting:
- 8F sheath and 260cm J inserted through guide delivery system.
- Guide delivery system removed, 20F dryseal inserted
- Agilis inserted
- Preload Shuttle sheath with glidewire and advance through agilis
- Glidewire used to cross between plugs and removed with shuttle dilator.
- 5F sheath inserted into shuttle in lieu of using torqvue delivery sideport
- Transition between shuttle and torqvue insertion isn't flush so occluder starts to deploy if sheath not used