TMVI - Encircle Sapien M3: Difference between revisions
From Bay Area Structural Heart Wiki
Line 13: | Line 13: | ||
*(1) 5F MICROPUNCTURE KIT | *(1) 5F MICROPUNCTURE KIT | ||
*(1) 6F | *(1) 6F SLENDER SHEATH | ||
*(1) 8F 10CM SHEATH | *(1) 8F 10CM SHEATH | ||
*(1) 8F 35CM BRITE TIP SHEATH | *(1) 8F 35CM BRITE TIP SHEATH | ||
Line 23: | Line 23: | ||
*(2) .035 150CM J WIRE | *(2) .035 150CM J WIRE | ||
*(1) .035 260CM J WIRE | *(1) .035 260CM J WIRE | ||
*(1) .032 260CM J WIRE | |||
*(1) SAFARI 2 WIRE | *(1) SAFARI 2 WIRE | ||
*(1) AMPLATZ SUPER STIFF 1CM TIP | *(1) AMPLATZ SUPER STIFF 1CM TIP | ||
Line 33: | Line 34: | ||
=='''Open'''== | =='''Open'''== | ||
*6F STRAIGHT | *(2) 6F STRAIGHT PIGTAIL 110CM | ||
*5F MPA1 | *5F MPA1 | ||
Line 55: | Line 55: | ||
*500ML HEPARINIZED SALINE | *500ML HEPARINIZED SALINE | ||
*(2) 60ML SYRINGE | *(2) 60ML SYRINGE | ||
*(2) 20ML SYRINGE | |||
*(1) HIGH PRESSURE STOPCOCK | *(1) HIGH PRESSURE STOPCOCK | ||
*30ML CONTRAST (MIX WITH 170ML HEP SALINE FOR 15% CONCENTRATION) | *30ML CONTRAST (MIX WITH 170ML HEP SALINE FOR 15% CONCENTRATION) | ||
=='''VALVE- SUPPLIED BY EDWARDS'''== | =='''VALVE- SUPPLIED BY EDWARDS'''== | ||
Line 92: | Line 91: | ||
*(2) PERCLOSE | *(2) PERCLOSE | ||
*3 PORT MANIFOLD | *3 PORT MANIFOLD | ||
*ACIST- | *ACIST- '''100% CONTRAST''' | ||
*PROBE COVER | *PROBE COVER | ||
*STAT LOCK WITH STRING SECUREMENT | *STAT LOCK WITH STRING SECUREMENT | ||
* | *(2) 0-Prolene | ||
*CHLORAPREP AND PREP TRAY | *CHLORAPREP AND PREP TRAY | ||
*DERMABOND | *DERMABOND | ||
*(2) FLUORO DRAPES (FOR LATERAL BASE AND IVY POLE ONCE IT IS MOVED TO THE END OF THE FIELD) | *(2) FLUORO DRAPES (FOR LATERAL BASE AND IVY POLE ONCE IT IS MOVED TO THE END OF THE FIELD) | ||
*36X36 BANDBAG (FOR LATERAL DETECTOR) | *36X36 BANDBAG (FOR LATERAL DETECTOR) | ||
*2L HEP SALINE ON | *2L HEP SALINE ON ALARIS PUMP | ||
*That's right, I said IVY pole. | *That's right, I said IVY pole. | ||
*Alaris pump | *(2) Alaris pump | ||
*Heparin tubing | *(2) Heparin tubing | ||
*72” pressure tubing | *(2) 72” pressure tubing | ||
*Swiftlink cover | *Swiftlink cover | ||
Line 158: | Line 157: | ||
*Edwards Guide Sheath (GS) advanced | *Edwards Guide Sheath (GS) advanced | ||
*DDS prepped | *DDS prepped | ||
**1L Hep. Saline bags with pressure bags spiked with heparin tubing, flushed, and attached to pump | **1L Hep. Saline (2U/2ml) bags with pressure bags spiked with heparin tubing, flushed, and attached to pump | ||
**Alaris settings 200ml/hr | **Alaris settings 200ml/hr occlusion pressure limit 300mm Hg | ||
**Pressure tubing attached to stopcock and DDS | **Pressure tubing attached to stopcock and DDS | ||
**Swiftlink cover installed over pressure tubing and secured ~18” from off-table end | **Swiftlink cover installed over pressure tubing and secured ~18” from off-table end |
Revision as of 23:46, 8 June 2020
DANIELS
Sapien M3
GENERAL ANESTHESIA, TEE, BIPLANE
ACCESS 2 FEMORAL VEINS (ESHEATH AND PACER), 1 FEMORAL ARTERY
TRIPLE LUMEN AND RADIAL ARTERY ACCESS BY ANESTHESIA
STABILIZER PLACEMENT PRIOR TO DRAPING
SHEATHS:
- (1) 5F MICROPUNCTURE KIT
- (1) 6F SLENDER SHEATH
- (1) 8F 10CM SHEATH
- (1) 8F 35CM BRITE TIP SHEATH
- (1) 8.5F 63CM SL1 TRANSSEPTAL SHEATH
WIRES:
Open
- (2) .035 150CM J WIRE
- (1) .035 260CM J WIRE
- (1) .032 260CM J WIRE
- (1) SAFARI 2 WIRE
- (1) AMPLATZ SUPER STIFF 1CM TIP
Standby
- Baylis ProTrack wire
CATHETERS
Open
- (2) 6F STRAIGHT PIGTAIL 110CM
- 5F MPA1
Standby
- 6F MPA1 125cm
PACING
- 5F JOSEPHSON PACING CATHETER
- EP QUAD CABLE
- DUAL RED/BLACK REMINGTON PACING CABLE
- (2) PACER BOXES
VALVE TABLE
- (5) STERILE BOWLS (3- 0.9% SALINE, 1-HEP SALINE, 1-15% CONTRAST MIX)
- (1) 90 INCH DRAPE
- 1.5L NORMAL SALINE
- 500ML HEPARINIZED SALINE
- (2) 60ML SYRINGE
- (2) 20ML SYRINGE
- (1) HIGH PRESSURE STOPCOCK
- 30ML CONTRAST (MIX WITH 170ML HEP SALINE FOR 15% CONCENTRATION)
VALVE- SUPPLIED BY EDWARDS
- INDEFLATOR
- CRIMPER
- E SHEATH
- DELIVERY SYSTEM
- VALVE
TRANS SEPTAL
- BAYLIS NEEDLE LARGE CURVE C1
- BAYLIS MACHINE
- BAYLIS CABLE
- GROUNDING PAD
SEPTAL DILATION
Open
- 14X4 ARMADA BALLOON- prepped
- PRESTO INDEFLATOR
Standby
- TORQUEVUE 45DEGREE DELIVERY SYSTEM (9F OR 10F)
- 10MM, 14MM, 18MM SEPTAL OCCLUDER
ADDITIONAL SUPPLIES
Open
- (2) PERCLOSE
- 3 PORT MANIFOLD
- ACIST- 100% CONTRAST
- PROBE COVER
- STAT LOCK WITH STRING SECUREMENT
- (2) 0-Prolene
- CHLORAPREP AND PREP TRAY
- DERMABOND
- (2) FLUORO DRAPES (FOR LATERAL BASE AND IVY POLE ONCE IT IS MOVED TO THE END OF THE FIELD)
- 36X36 BANDBAG (FOR LATERAL DETECTOR)
- 2L HEP SALINE ON ALARIS PUMP
- That's right, I said IVY pole.
- (2) Alaris pump
- (2) Heparin tubing
- (2) 72” pressure tubing
- Swiftlink cover
MEDICATIONS
- BUPIVACAINE/LIDOCAINE MIX
- EPINEPHRINE 10MCG/ML IN RED SYRINGE ON BACK TABLE
- ROTAGLIDE 10ML (PRE VALVE)
STANDBY EMERGENCY ECMO SUPPLIES
- 16F ARTERIAL CANNULA
- 21F VENOUS CANNULA
- (2) PERCLOSE
- ECMO INSTRUMENT TRAY
- AMPLATZ SUPER STIFF
STANDBY PERICARDIOCENTESIS SUPPLIES
- EYE DRAPE
- CHIBA NEEDLE 10CM 18G
- SKATER INTRODUCER SET
- BARD LOCKING PIGTAIL 6F OR 8F
- (3) 60ML SYRINGES
- STOPCOCK
- AMPLATZ SUPER STIFF 180CM
STANDBY PARAVALVULAR LEAK REPAIR SUPPLIES
SEE MITRAL PARAVALVULAR LEAK REPAIR
- AMPLATZER VASCULAR PLUG 2 SIZE 8-18 AVAILABLE
PROCEDURE
- Patient prepped from navel to knees in sterile fashion.
- ISOCENTER XRAY
- Access obtained with u/s
Typical sites:
- 6F LFA for Pigtail
- 8F 35cm LFV for transcutaneous pacing
- 20F RFV E-sheath for valve delivery
- Dilate with 8F
- Preclose
- Pigtail advanced through 6F sheath and LVgram performed
- Mitral annulus denoted on live monitor screen in A plane
- SL1 inserted over .032 kit wire
- Baylis needle inserted through SL1 and atrial septum puncture performed.
- Needle removed
- 5F MPA advanced over .032 wire
- 1cm Amplatz Super Stiff advanced and MPA and SL1 removed
- Edwards Guide Sheath (GS) advanced
- DDS prepped
- 1L Hep. Saline (2U/2ml) bags with pressure bags spiked with heparin tubing, flushed, and attached to pump
- Alaris settings 200ml/hr occlusion pressure limit 300mm Hg
- Pressure tubing attached to stopcock and DDS
- Swiftlink cover installed over pressure tubing and secured ~18” from off-table end
- After hookup and flush, Swiftlink cover can be slid over heparin tubing for additional slack on the table.
- Dock Delivery System (DDS) advanced through GS
- Dock deployed in a timely fashion ;)
- DDS removed and Pigtail inserted over exchange J for valve crossing
- J wire exchanged for Safari 2 and Pigtail removed
- 14mm balloon advanced for septostomy
- Balloon and GS removed
- Valve delivery system advanced
- When valve is in place, temp pacer inserted and tested.
- Valve deployed. Septum evaluated with echo for shunting and PVL.
- Delivery system removed and MPA inserted for LA pressure measurement.
- All equipment removed
- Preclosure closed and stitch to RFV
- Perclose LFA
- 8F vein exchanged for 9F 10cm sheath
- Secure sheath with Statlock. Tegaderm both sides.
APPROVED: MD initials MM/YY