==Equipment==
==Equipment==
*Atrion 38 indeflator
*Atrion 38 indeflator
*Atrion 25 indeflator
*Atrion 25 indeflator
*4-way stopcock
*4-way stopcock
*True Balloon
*True Balloon
**23mm - 30cc in big indeflator, 10cc in small indeflator
**Fill big indeflator, 10ml in small indeflator


==Procedure==
==Procedure==
#Use 30/170 TAVR mix
#Use 30/170 TAVR mix
#De-air and fill both indeflators
#De-air and fill both indeflators
Line 15: Line 17:
#Attach both indeflators and open stopcock to be open all ways
#Attach both indeflators and open stopcock to be open all ways
#Lock smaller indeflator
#Lock smaller indeflator
#Pull negative and lock larger indeflator
#Leave larger indeflator unlocked


[[Category:Setups]]
[[Category:Setups]]
[[Category:Crazy Gems]]

Latest revision as of 21:25, 10 December 2021

Valve Cracker Setup

This setup is used to over dilate an existing valve in preparation for installing a new trans catheter valve.

Equipment

  • Atrion 38 indeflator
  • Atrion 25 indeflator
  • 4-way stopcock
  • True Balloon
    • Fill big indeflator, 10ml in small indeflator

Procedure

  1. Use 30/170 TAVR mix
  2. De-air and fill both indeflators
  3. Prep balloon similar to Edwards Delivery system prep
  4. Attach both indeflators and open stopcock to be open all ways
  5. Lock smaller indeflator
  6. Leave larger indeflator unlocked