Maclab FFR/DFR: Difference between revisions
From Bay Area Structural Heart Wiki
Created page with "Step 1 - Turn on P2 channel by either double clicking P2, or clicking P2 on keyboard thumb|218x218px Step 2 - Open REAL TIME tab, then select FFR display on. This step will not work if you have not opened P2 yet. thumb|223x223px Step 3 - Once wire is inserted and physician is ready, equalize by double clicking in FFR square. After equalized switch to DFR if not using adenosine. '''THIS MAY TAKE A F..." |
No edit summary |
||
(3 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
__NOTOC__ | |||
==Step 1== | |||
*Turn on P2 channel by either double clicking P2, or clicking P2 on keyboard | |||
[[File:FFR Step 1.jpg|left|thumb|218x218px]] | |||
<div style="clear: both"></div> | |||
==Step 2== | |||
*Open REAL TIME tab, then select FFR display on. This step will not work if you have not opened P2 yet. | |||
[[File:FFR step 2.jpg|left|thumb|223x223px]] | |||
<div style="clear: both"></div> | |||
==Step 3== | |||
*Once wire is inserted and physician is ready, equalize by double clicking in FFR square. | |||
*After equalized switch to DFR if not using adenosine. | |||
'''THIS MAY TAKE A FEW ATTEMPTS FOR PRESSURE TO EQUALIZE, THIS IS NORMAL!!!!!''' | |||
Step | ==Step 4== | ||
*Once equalized and on desired measurement selection (FFR or DFR), open MEASUREMENTS tab and select FFR/DFR | |||
[[File:FFR Step 3.jpg|left|thumb|269x269px]] | |||
<div style="clear: both"></div> | |||
==Step 5== | |||
*A window will open on the RIGHT sided monitor, all you have to do is click measure, you do NOT have to select a vessel. | |||
*The measurement should populate under baseline measurements after a few beats. | |||
[[Category:Setups]] | |||
Latest revision as of 21:05, 26 February 2025
Step 1
- Turn on P2 channel by either double clicking P2, or clicking P2 on keyboard

Step 2
- Open REAL TIME tab, then select FFR display on. This step will not work if you have not opened P2 yet.

Step 3
- Once wire is inserted and physician is ready, equalize by double clicking in FFR square.
- After equalized switch to DFR if not using adenosine.
THIS MAY TAKE A FEW ATTEMPTS FOR PRESSURE TO EQUALIZE, THIS IS NORMAL!!!!!
Step 4
- Once equalized and on desired measurement selection (FFR or DFR), open MEASUREMENTS tab and select FFR/DFR

Step 5
- A window will open on the RIGHT sided monitor, all you have to do is click measure, you do NOT have to select a vessel.
- The measurement should populate under baseline measurements after a few beats.