Laser CVX-300: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
|||
(3 intermediate revisions by the same user not shown) | |||
Line 6: | Line 6: | ||
==OPERATION== | ==OPERATION== | ||
*Insert Catheter and Calibrate | *Insert Catheter and Calibrate | ||
*Depress the Calibrate button, aim the catheter and 1-2 inches away from the energy detector | **Depress the Calibrate button, aim the catheter and 1-2 inches away from the energy detector | ||
*Depress the foot switch until the system completes calibration approximately 5 seconds) -The "Calibration Complete" or Cal OK" lamp is displayed | **Depress the foot switch until the system completes calibration approximately 5 seconds) -The "Calibration Complete" or Cal OK" lamp is displayed | ||
*Set the Fluence and Rate Per Physician's Instructions (i.e. 50/50) Select the "Ready" Mode | *Set the Fluence and Rate Per Physician's Instructions (i.e. 50/50) | ||
*Depress foot pedal per Physician's instructions until lesion(s) modification is complete. | *Select the "Ready" Mode | ||
**Depress foot pedal per Physician's instructions until lesion(s) modification is complete. | |||
*Select "Standby" Mode when the laser i or is removed from body. | *Select "Standby" Mode when the laser i or is removed from body. | ||
Line 15: | Line 16: | ||
==REPORTING== | ==REPORTING== | ||
*Record "Pulses Delivered" in log book | *Record "Pulses Delivered" in log book | ||
*Depressing the "Pulses Delivered" button enables the total number of pulses during a procedure to be visible in the display window Record "Treatment Time" in log book | **Depressing the "Pulses Delivered" button enables the total number of pulses during a procedure to be visible in the display window | ||
*Depressing the "Treatment Time" button enables the total lasing time of the procedure to be visible in the display window | *Record "Treatment Time" in log book | ||
**Depressing the "Treatment Time" button enables the total lasing time of the procedure to be visible in the display window | |||
==TROUBLESHOOTING== | ==TROUBLESHOOTING== | ||
Line 31: | Line 33: | ||
==SHUTDOWN/STORAGE== | ==SHUTDOWN/STORAGE== | ||
*Turn the Laser System OFF | *Turn the Laser System OFF | ||
*Disconnect and store the power cord | **Disconnect and store the power cord | ||
*Disconnect and store the footswitch | **Disconnect and store the footswitch | ||
*Close the catheter connector door | **Close the catheter connector door | ||
*Clean the energy detector's face with alcohol prep | **Clean the energy detector's face with alcohol prep | ||
*Cover the laser system with the dust cover | **Cover the laser system with the dust cover | ||
[[Category:Products]] | |||
[[Category:IRProducts]] |
Latest revision as of 17:37, 7 November 2024
PHILIPS CVX-300 QUICK-START CHECKLIST
SET-UP
- Plug in CVX-300 and Turn Power Switch ON System will Complete Self-Test (5 minute warm up) (An audible beep will sound indicating the laser system is ready)
- Connect Foot Pedal
OPERATION
- Insert Catheter and Calibrate
- Depress the Calibrate button, aim the catheter and 1-2 inches away from the energy detector
- Depress the foot switch until the system completes calibration approximately 5 seconds) -The "Calibration Complete" or Cal OK" lamp is displayed
- Set the Fluence and Rate Per Physician's Instructions (i.e. 50/50)
- Select the "Ready" Mode
- Depress foot pedal per Physician's instructions until lesion(s) modification is complete.
- Select "Standby" Mode when the laser i or is removed from body.
After Treatment is Complete
REPORTING
- Record "Pulses Delivered" in log book
- Depressing the "Pulses Delivered" button enables the total number of pulses during a procedure to be visible in the display window
- Record "Treatment Time" in log book
- Depressing the "Treatment Time" button enables the total lasing time of the procedure to be visible in the display window
TROUBLESHOOTING
- Laser System Will Not Turn ON
- Verify power cord connections
- Verify interlock plug installed
- Verify main circuit breaker is in ON position
- Laser Buzzes When Plugged In
- Release red emergency button
- Fault Codes 1-4, "Service", "Service Laser Soon", "Power Error"
- Shut machine OFF, wait 30 seconds, then Turn ON
- Call PHILIPS Customer Service @ 800-231-0978 for assistance
SHUTDOWN/STORAGE
- Turn the Laser System OFF
- Disconnect and store the power cord
- Disconnect and store the footswitch
- Close the catheter connector door
- Clean the energy detector's face with alcohol prep
- Cover the laser system with the dust cover