Rotablator: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
Line 22: Line 22:
#Open nitrogen tank valve.
#Open nitrogen tank valve.
#Note tank pressure. (If below 1000, consider replacement)
#Note tank pressure. (If below 1000, consider replacement)
#*BSC guidelines are minimum of 500 psi for starting a case!
#Adjust regulator to appropriate pressure range (90-110 psi) using adjustment valve.
#Adjust regulator to appropriate pressure range (90-110 psi) using adjustment valve.
#*Adjust slowly using arrows on knob as a guide.
#*Adjust slowly using arrows on knob as a guide.

Revision as of 18:39, 7 April 2020

aka RotaPro

Manufacturer: Boston Scientific

Product website: BSC RotaPro


Setup

Setup Video

The RotaPro is kept behind Room 6. You need the IV pole with the device, the nitrogen tank, 108” tubing, and the appropriate burr.

Open the burr and tubing for the scrub person to setup.

Rota Burr Location
Rota tank valves
Rota tank valve detail
RotaPro Console
Tank Wrench Location

Circulator

  1. Attach air hose to back of device.
  2. Open nitrogen tank valve.
  3. Note tank pressure. (If below 1000, consider replacement)
    • BSC guidelines are minimum of 500 psi for starting a case!
  4. Adjust regulator to appropriate pressure range (90-110 psi) using adjustment valve.
    • Adjust slowly using arrows on knob as a guide.
    • Pressure will not go down as you turn it down unless you vent the hose.
    • Vent hose by partially removing it from regulator.
    • If pressure isn’t increasing, check the tank valve.
  5. Plug in power cord to console
    • If it’s not turning on, check main power switch on back.
  6. Connect umbilical from sterile device to front.
  7. Use front pressure knob to adjust speed to MD’s desired range.

Scrub

  1. Turn package on its lid and open on table. (The lid is useful as a fluid container)
  2. Pass off umbilical to circulator.
  3. Pass off IV spike for tubing. (Keep roller clamp)
  4. Attach tubing to short tubing piece on device.
  5. Once Rotaglide is flowing, release and slide advancer to exercise it and clear bubbles.
  6. Lock advancer ~2 cm from fully retracted.
  7. Load on wire.
  8. Once wire is out the back, place clip on end of wire.
  9. Observe rotaglide fluid is flowing from tip of device.
  10. Activate device outside the body when physician is ready to test and adjust speed.
  11. Rotablate!

Breakdown

  1. Disconnect sterile device umbilical.
  2. Unplug power.
  3. Close main tank valve
  4. Carefully unplug air hose from console.

General Notes

  • Ensure tank valve is closed!
  • Tank wrench is located on base of tank.