Cryo Setup: Difference between revisions

From Bay Area Structural Heart Wiki
(Created page with "Cryo page")
 
(in progress)
 
Line 1: Line 1:
Cryo page
<u>'''Cryo Equipment/Supplies'''</u>
 
* Cryo Console (Located in sterile core)
* Cryo Catheter (4,6,or 8mm)
* Electrical Umbilical (Sterile core)
* Coaxial Umbilical  (Sterile core)
* 8F sheath, usually a long sheath like an SR0
 
'''<u>Steps</u>'''
 
* Get cryo console and position towards the foot of table on scrub side
* Connect Scavenging Hose (white hose on console) to vacuum port on boom near anesthesia
* Plug in power cable
* Open refrigerant tank in console (Lefty Loosey)
* Power up console (It is recommended that power is on at least 15mins prior to starting procedure)
* Check for error messages (If yes, check all connections and refrigerant tank is open and full. Reboot)
* After power up - Complete patient info screen
* Connect Auto-connection box (located inside the console) to the front of the console
* Plug pin connector box (located in room at bottom of 3rd cabinet) into 20pole B
* Connect Auto connection box pins in 20pole B 1 through 4 (Distal to Proximal)
* Connect Electrical Umbilical to Auto connection box (From sterile table to ACB)
* Connect Coaxial Umbilical to front of Console (From Sterile table to Console)
* Connect Catheter to Electrical and coaxial umbilicals
* On Prucka, go to Pg 8
 
<nowiki>*</nowiki>
 
 
 
<br />

Latest revision as of 21:13, 11 February 2020

Cryo Equipment/Supplies

  • Cryo Console (Located in sterile core)
  • Cryo Catheter (4,6,or 8mm)
  • Electrical Umbilical (Sterile core)
  • Coaxial Umbilical (Sterile core)
  • 8F sheath, usually a long sheath like an SR0

Steps

  • Get cryo console and position towards the foot of table on scrub side
  • Connect Scavenging Hose (white hose on console) to vacuum port on boom near anesthesia
  • Plug in power cable
  • Open refrigerant tank in console (Lefty Loosey)
  • Power up console (It is recommended that power is on at least 15mins prior to starting procedure)
  • Check for error messages (If yes, check all connections and refrigerant tank is open and full. Reboot)
  • After power up - Complete patient info screen
  • Connect Auto-connection box (located inside the console) to the front of the console
  • Plug pin connector box (located in room at bottom of 3rd cabinet) into 20pole B
  • Connect Auto connection box pins in 20pole B 1 through 4 (Distal to Proximal)
  • Connect Electrical Umbilical to Auto connection box (From sterile table to ACB)
  • Connect Coaxial Umbilical to front of Console (From Sterile table to Console)
  • Connect Catheter to Electrical and coaxial umbilicals
  • On Prucka, go to Pg 8

*