From Bay Area Structural Heart Wiki
(Created page with "This setup is used to over dilate an existing valve in preparation for installing a new trans catheter valve. ==Equipment== *Atrion 38 indeflator *Atrion 25 indeflator *4-way...") |
No edit summary |
||
Line 1: | Line 1: | ||
[[File:Valve_Cracker.jpg|thumb|150px|right|Valve Cracker Setup|link={{filepath:Valve_Cracker.jpg}}]] | |||
This setup is used to over dilate an existing valve in preparation for installing a new trans catheter valve. | This setup is used to over dilate an existing valve in preparation for installing a new trans catheter valve. | ||
Revision as of 21:53, 9 December 2019
This setup is used to over dilate an existing valve in preparation for installing a new trans catheter valve.
Equipment
- Atrion 38 indeflator
- Atrion 25 indeflator
- 4-way stopcock
- True Balloon
- 23mm - 30cc in big indeflator, 10cc in small indeflator
Procedure
- Use 30/170 TAVR mix
- De-air and fill both indeflators
- Prep balloon similar to Edwards Delivery system prep
- Attach both indeflators and open stopcock to be open all ways
- Lock smaller indeflator
- Pull negative and lock larger indeflator