*14frx33cm Gore Dry Seal Flex
*14frx33cm Gore Dry Seal Flex
*6fr MP1 guide catheter  
*6fr MP1 guide catheter
*.018 Platinum Plus guidewire 260 cm
*.018 Platinum Plus guidewire 260 cm
*Gooseneck Snare 20mm
*Gooseneck Snare 20mm
Line 23: Line 23:
*Gooseneck snares 25, 30mm
*Gooseneck snares 25, 30mm
*Piggyback wire converter 4fr x 145 (On standby)
*Piggyback wire converter 4fr x 145 (On standby)
*.035 QuickCross (On standby)  
*.035 QuickCross (On standby)
*AL2, AL3, EBU4, EBU 3.75 (Left leaflet)
*AL2, AL3, EBU4, EBU 3.75 (Left leaflet)
*PAR1, JR4, IM, AR1 (Right leaflet)
*PAR1, JR4, IM, AR1 (Right leaflet)
Line 29: Line 29:
*(2) .035 torque devices
*(2) .035 torque devices
*Finecross 135
*Finecross 135
<br />
=== Procedure ===
* Setup and access is the same as TAVR '''''except''''' a 14F Dryseal sheath is inserted initially instead of the ESheath.
** Bovie set at 30 CUT ONLY.
* AL1 advanced over J wire through 6F sheath.  Wire exchanged for straight stiff glidewire for valve crossing.
* AL1 exchanged over the wire for MP1 Guide with watchdog and stopcock attached.
* Platinum Plus wire is preshaped by MD and inserted through MP to retain LV access.
* Snare advanced through MP alongside Platinum Plus.
* BASILICA catheter advanced through Dryseal sheath with watchdog and stopcock attached.
* Kinked Astato wire advanced through BASILICA catheter.
* Hand injection performed to verify position of BASILICA catheter.
* Astato wire connected to bovie with 2 hemostats.
* RF applied as Astato wire advanced.
* Once wire penetrates leaflet it is snared and pulled into MP1. 
* Platinum Plus wire removed
* Pacing wire advanced and tested
* Al1 advanced next to BASILICA catheter and valve recrossed in standard fashion.
* Leaflet lacerated
* MP removed. BASILICA catheter removed.
* Amplatz ES advanced through AL1 and AL1 removed.
* Pigtail advanced to coronary cusp
* Dryseal exchanged for ESheath
* Do a TAVR.





Revision as of 00:06, 26 October 2019

(Bioprosthetic Aortic Scallop Intentional Laceration to prevent Iatrogenic Coronary Artery obstruction)

ACC.org BASILICA Page

Open

  • 14frx33cm Gore Dry Seal Flex
  • 6fr MP1 guide catheter
  • .018 Platinum Plus guidewire 260 cm
  • Gooseneck Snare 20mm
  • 6 Fr PAL 1.0 (for LCA BASILICA) -OR- 6 FR MPA1 Guide (For RCA BASILICA)
  • .014 Astato xs 20 guidewire 300 cm
  • Bovie pen
    • Grounding Pad
    • Bovie machine
  • (2) Watchdog devices
  • (2) Stopcocks
  • 4 hemostats

Standby

  • Gooseneck snares 25, 30mm
  • Piggyback wire converter 4fr x 145 (On standby)
  • .035 QuickCross (On standby)
  • AL2, AL3, EBU4, EBU 3.75 (Left leaflet)
  • PAR1, JR4, IM, AR1 (Right leaflet)
  • (3) .014 torque devices
  • (2) .035 torque devices
  • Finecross 135


Procedure

  • Setup and access is the same as TAVR except a 14F Dryseal sheath is inserted initially instead of the ESheath.
    • Bovie set at 30 CUT ONLY.
  • AL1 advanced over J wire through 6F sheath. Wire exchanged for straight stiff glidewire for valve crossing.
  • AL1 exchanged over the wire for MP1 Guide with watchdog and stopcock attached.
  • Platinum Plus wire is preshaped by MD and inserted through MP to retain LV access.
  • Snare advanced through MP alongside Platinum Plus.
  • BASILICA catheter advanced through Dryseal sheath with watchdog and stopcock attached.
  • Kinked Astato wire advanced through BASILICA catheter.
  • Hand injection performed to verify position of BASILICA catheter.
  • Astato wire connected to bovie with 2 hemostats.
  • RF applied as Astato wire advanced.
  • Once wire penetrates leaflet it is snared and pulled into MP1.
  • Platinum Plus wire removed
  • Pacing wire advanced and tested
  • Al1 advanced next to BASILICA catheter and valve recrossed in standard fashion.
  • Leaflet lacerated
  • MP removed. BASILICA catheter removed.
  • Amplatz ES advanced through AL1 and AL1 removed.
  • Pigtail advanced to coronary cusp
  • Dryseal exchanged for ESheath
  • Do a TAVR.




Approved: MM/YY