Watchman Retrieval: Difference between revisions

From Bay Area Structural Heart Wiki
Rosenbzj (talk | contribs)
Created page with "To Drop: preclose straight 6 pig 33cm 20f gore (2cc saline to prep valve) Grandslam 300cm 260cm 0.035" J Amplatz 260cm 1cm floppy tip VersaCross 260cm TrueSteer (poss 13f Agilis instead, Ono thru Truesteer was tight though case was successful) Raptor Ono 50/60cc syringe 1L of hep saline on the table manifold + pressure line sentinel 5/6 slender radial sheath 2.5 of verapamil (instead of slurry for Spies) Tegaderm for radial sheath Radial drape tr band Procedural or..."
 
Rosenbzj (talk | contribs)
No edit summary
 
(4 intermediate revisions by the same user not shown)
Line 1: Line 1:
To Drop:
set fluoro to card alt 10p/s
 
drugs - verapamil, lido+epi, 1L of hep saline for the table, normal manifold setup
 
 
 
 
'''To Drop:'''


preclose
preclose
straight 6 pig
straight 6 pig
33cm 20f gore (2cc saline to prep valve)
33cm 20f gore (2cc saline to prep valve)
Grandslam 300cm
Grandslam 300cm
260cm 0.035" J
260cm 0.035" J
Amplatz 260cm 1cm floppy tip
Amplatz 260cm 1cm floppy tip
VersaCross 260cm
VersaCross 260cm
TrueSteer (poss 13f Agilis instead, Ono thru Truesteer was tight though case was successful)
TrueSteer (poss 13f Agilis instead, Ono thru Truesteer was tight though case was successful)
Raptor
Raptor
Ono
Ono
50/60cc syringe
50/60cc syringe
1L of hep saline on the table
1L of hep saline on the table
manifold + pressure line
manifold + pressure line
sentinel
sentinel
---------------------


5/6 slender radial sheath
5/6 slender radial sheath
2.5 of verapamil (instead of slurry for Spies)
2.5 of verapamil (instead of slurry for Spies)
Tegaderm for radial sheath
Tegaderm for radial sheath
Radial drape  
Radial drape  
tr band
tr band






Procedural order:


1 radial access
 
2 verapamil
'''Procedural order:'''
3 pigtail + wire to arch
 
4 arch picture thru pigtail (hand inject w/manifold)
1) radial access
5 grandslam
 
6 sentinel
2) verapamil
7 rfv micro access + preclose + dilate
 
8 gore 20f (2cc saline in valve to prep) + versacross telescoping to transeptal
3) pigtail + wire to arch
9 upsize versacross to either truesteer or agilis 13f (truesteer was tight interacting w/ono last case)
 
10 ono
4) arch picture thru pigtail (hand inject w/manifold)
11 raptor
 
12 retrieve
5) grandslam
13 arch shot post sentinel removal
 
6) sentinel
 
7) rfv micro access + preclose + dilate
 
8) gore 20f (2cc saline in valve to prep) + versacross telescoping to transeptal
 
9) upsize versacross to either truesteer or agilis 13f (truesteer was tight interacting w/ono last case)
 
10) ono
 
11) raptor
 
12) retrieve
 
13) arch shot post sentinel removal
 




poss reimplant
poss reimplant


tr band (12cc, green dot on sheath entry site) /groin closure
tr band (12cc, green dot on sheath entry site) /groin closure

Latest revision as of 01:03, 16 December 2025

set fluoro to card alt 10p/s

drugs - verapamil, lido+epi, 1L of hep saline for the table, normal manifold setup



To Drop:

preclose

straight 6 pig

33cm 20f gore (2cc saline to prep valve)

Grandslam 300cm

260cm 0.035" J

Amplatz 260cm 1cm floppy tip

VersaCross 260cm

TrueSteer (poss 13f Agilis instead, Ono thru Truesteer was tight though case was successful)

Raptor

Ono

50/60cc syringe

1L of hep saline on the table

manifold + pressure line

sentinel


5/6 slender radial sheath

2.5 of verapamil (instead of slurry for Spies)

Tegaderm for radial sheath

Radial drape

tr band



Procedural order:

1) radial access

2) verapamil

3) pigtail + wire to arch

4) arch picture thru pigtail (hand inject w/manifold)

5) grandslam

6) sentinel

7) rfv micro access + preclose + dilate

8) gore 20f (2cc saline in valve to prep) + versacross telescoping to transeptal

9) upsize versacross to either truesteer or agilis 13f (truesteer was tight interacting w/ono last case)

10) ono

11) raptor

12) retrieve

13) arch shot post sentinel removal


poss reimplant


tr band (12cc, green dot on sheath entry site) /groin closure