Woods Farapulse: Difference between revisions
Created page with "work in progress General: -Consent - Electrophysiology Study with Catheter Ablation -Patient intubated/paralyzed (long acting) throughout procedure -No foley -No isuprel -Antibiotics per woods protocol (only w/device, implants etc.) -Fluoro -Goal ACT>300 -McKesson leads, Anesthesia leads, Farapulse leads (slaved into Claris 12 lead by rep) -2 Grounds, Abbott patches, regular defib pad From Pyxis: -protamine if patient is going home. no protamine if patient is..." |
No edit summary |
||
| (18 intermediate revisions by 2 users not shown) | |||
| Line 1: | Line 1: | ||
General: | General: | ||
| Line 16: | Line 16: | ||
-Goal ACT>300 | -Goal ACT>300 | ||
- | -MacLab leads, Anesthesia leads, Claris 12 lead | ||
- | -1 ground, Abbott patches, regular defib pad | ||
From Pyxis: | From Pyxis: | ||
-protamine | -Case by case basis for protamine | ||
-famotidine | -No famotidine | ||
-bupivicaine | -bupivicaine | ||
| Line 32: | Line 32: | ||
- (1x) heparin premixed (25,000/250mL) | - (1x) heparin premixed (25,000/250mL) | ||
- for typical AFL PFA: Nitro - 250ml bottle of 200mcg/ml | |||
| Line 44: | Line 47: | ||
-(1x) 500mL NS with 5000u heparin for the table | -(1x) 500mL NS with 5000u heparin for the table | ||
-( | -(2x) 500mL NS with 5000u heparin for red/yellow lines | ||
-(1x prepulled from pyxis) heparin drip for blue line | -(1x prepulled from pyxis 25,000u heparin/250mL NS) heparin drip for blue line | ||
Lines: | Lines: | ||
| Line 54: | Line 57: | ||
-Blue (heparin drip) | -Blue (heparin drip) | ||
-Red ( | -Red (Grid - Heparin 5000u/500mL NS @ 120cc/hr) | ||
-Yellow (Farapulse - Heparin 5000u/500mL NS run at 60-120cc/hr) | |||
Charting ablation in Cupid: | Charting ablation in Cupid: | ||
System used - freetext Farapulse | System used - freetext Farapulse | ||
Energy - freetext PFA | Energy - freetext PFA | ||
Number of energy applications - # per Ablation site | Number of energy applications - # per Ablation site | ||
| Line 69: | Line 80: | ||
8F pinnacle (ICE) | 8F pinnacle (ICE) | ||
8f locking sheath (Inquiry | 8f locking sheath (DuoDeca or Inquiry) | ||
Dilator from 8f pinnacle (---> Preclose ----> VersaCrosswire ---> 13F Agilis ---> Map with Grid -----> Swap Grid for Farawave & Ablate) | |||
| Line 79: | Line 90: | ||
Rampart + Radpad | Rampart + Radpad | ||
3x yellow filters | |||
8f Acunav (non-soundstar) | 8f Acunav (non-soundstar) | ||
1x Perclose | |||
DuoDeca + Cable | |||
Grid-X + Cable | |||
Light handle cover | Light handle cover | ||
| Line 93: | Line 104: | ||
VersaCross 180cm | |||
Micropuncture | |||
1x Ice Sleeve | 1x Ice Sleeve | ||
1x Access sleeve | |||
ACT line | ACT line | ||
3x pressure lines (label blue/red/yellow) | |||
Ioban | Ioban | ||
8F pinnacle | 8F pinnacle x2 | ||
8f locking sheath | 8f locking sheath | ||
ask - driver/0pds/stopcock for woggle if no protamine | |||
driver/0pds/stopcock for woggle | |||
chloraprep | chloraprep | ||
vascade | vascade x2 | ||
Rosen wire | Rosen wire w/1.5mmJ | ||
[[Category:Ablations]] | |||
Latest revision as of 23:18, 5 November 2025
General:
-Consent - Electrophysiology Study with Catheter Ablation
-Patient intubated/paralyzed (long acting) throughout procedure
-No foley
-No isuprel
-Antibiotics per woods protocol (only w/device, implants etc.)
-Fluoro
-Goal ACT>300
-MacLab leads, Anesthesia leads, Claris 12 lead
-1 ground, Abbott patches, regular defib pad
From Pyxis:
-Case by case basis for protamine
-No famotidine
-bupivicaine
- (2x) heparin vials (10,000 u/10mL)
- (1x) heparin premixed (25,000/250mL)
- for typical AFL PFA: Nitro - 250ml bottle of 200mcg/ml
Bags:
-500mL NS for ACT line
-1L LR anesthesia
-(1x) 500mL NS with 5000u heparin for the table
-(2x) 500mL NS with 5000u heparin for red/yellow lines
-(1x prepulled from pyxis 25,000u heparin/250mL NS) heparin drip for blue line
Lines:
-ACT
-Blue (heparin drip)
-Red (Grid - Heparin 5000u/500mL NS @ 120cc/hr)
-Yellow (Farapulse - Heparin 5000u/500mL NS run at 60-120cc/hr)
Charting ablation in Cupid:
System used - freetext Farapulse
Energy - freetext PFA
Number of energy applications - # per Ablation site
Drop List:
Sheaths:
8F pinnacle (ICE)
8f locking sheath (DuoDeca or Inquiry)
Dilator from 8f pinnacle (---> Preclose ----> VersaCrosswire ---> 13F Agilis ---> Map with Grid -----> Swap Grid for Farawave & Ablate)
Rampart + Radpad
3x yellow filters
8f Acunav (non-soundstar)
1x Perclose
DuoDeca + Cable
Grid-X + Cable
Light handle cover
VersaCross 180cm
Micropuncture
1x Ice Sleeve
1x Access sleeve
ACT line
3x pressure lines (label blue/red/yellow)
Ioban
8F pinnacle x2
8f locking sheath
ask - driver/0pds/stopcock for woggle if no protamine
chloraprep
vascade x2
Rosen wire w/1.5mmJ