Double Pump Valve Cracker: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
 
(3 intermediate revisions by 2 users not shown)
Line 4: Line 4:
==Equipment==
==Equipment==


*Atrion 38 indeflator
* Atrion 38 indeflator
*Atrion 25 indeflator
* Atrion 25 indeflator
*4-way stopcock
* 4-way stopcock
*True Balloon
* True Balloon (1mm smaller than intended Sapien implant)
**23mm - 30cc in big indeflator, 10-15cc in small indeflator
** Fill big indeflator '''all the way''', 20ml in small indeflator


==Procedure==
==Procedure==


#Use 30/170 TAVR mix
# Use TAVR mix
#De-air and fill both indeflators
# De-air and fill both indeflators
#Prep balloon similar to Edwards Delivery system prep
# Prep balloon similar to Edwards Delivery system prep
#Attach both indeflators and open stopcock to be open all ways
# Attach both indeflators and open stopcock to be open all ways
#Lock smaller indeflator
# Lock both indeflators
#Leave larger indeflator unlocked


[[Category:Setups]]
[[Category:Setups]]
[[Category:Crazy Gems]]

Latest revision as of 18:24, 12 March 2025

Valve Cracker Setup

This setup is used to over dilate an existing valve in preparation for installing a new trans catheter valve.

Equipment

  • Atrion 38 indeflator
  • Atrion 25 indeflator
  • 4-way stopcock
  • True Balloon (1mm smaller than intended Sapien implant)
    • Fill big indeflator all the way, 20ml in small indeflator

Procedure

  1. Use TAVR mix
  2. De-air and fill both indeflators
  3. Prep balloon similar to Edwards Delivery system prep
  4. Attach both indeflators and open stopcock to be open all ways
  5. Lock both indeflators