==Equipment== | ==Equipment== | ||
*Atrion 38 indeflator | * Atrion 38 indeflator | ||
*Atrion 25 indeflator | * Atrion 25 indeflator | ||
*4-way stopcock | * 4-way stopcock | ||
*True Balloon | * True Balloon (1mm smaller than intended Sapien implant) | ||
** | ** Fill big indeflator '''all the way''', 20ml in small indeflator | ||
==Procedure== | ==Procedure== | ||
#Use | # Use TAVR mix | ||
#De-air and fill both indeflators | # De-air and fill both indeflators | ||
#Prep balloon similar to Edwards Delivery system prep | # Prep balloon similar to Edwards Delivery system prep | ||
#Attach both indeflators and open stopcock to be open all ways | # Attach both indeflators and open stopcock to be open all ways | ||
#Lock | # Lock both indeflators | ||
[[Category:Setups]] | [[Category:Setups]] | ||
[[Category:Crazy Gems]] |
Latest revision as of 18:24, 12 March 2025

This setup is used to over dilate an existing valve in preparation for installing a new trans catheter valve.
Equipment
- Atrion 38 indeflator
- Atrion 25 indeflator
- 4-way stopcock
- True Balloon (1mm smaller than intended Sapien implant)
- Fill big indeflator all the way, 20ml in small indeflator
Procedure
- Use TAVR mix
- De-air and fill both indeflators
- Prep balloon similar to Edwards Delivery system prep
- Attach both indeflators and open stopcock to be open all ways
- Lock both indeflators