Created page with "<nowiki>***</nowiki> THIS IS A RESEARCH CASE ; EQUIPMENT WILL NOT SCAN, PLEASE KEEP BARCODES OF ALL NON-SCANNABLE EQUIPMENT FOR ROSALIE WITH A PATIENT LABEL *** As of today 1/25/2025 we have done ONE of these procedures ( on 3/29/2024 ). This page is rough draft and will be updated along the way. '''Room Set Up ; 2 Manifolds needed for pressure only. No Acist. IV Pole for back table similarly to MitraClip with 1L pressure bag. Will need the MitraClip plate and stool..."
(No difference)

Revision as of 17:13, 15 January 2025

*** THIS IS A RESEARCH CASE ; EQUIPMENT WILL NOT SCAN, PLEASE KEEP BARCODES OF ALL NON-SCANNABLE EQUIPMENT FOR ROSALIE WITH A PATIENT LABEL ***

As of today 1/25/2025 we have done ONE of these procedures ( on 3/29/2024 ). This page is rough draft and will be updated along the way.

Room Set Up ; 2 Manifolds needed for pressure only. No Acist. IV Pole for back table similarly to MitraClip with 1L pressure bag. Will need the MitraClip plate and stool over the patient; distance from zyphoid unknown...Radial prepped for pigtail. Rampart, yes if the deployment angles allow.

Supplies

  • TAVI Pack
  • Micropuncture
  • 8Fr Pinnacle
  • 2 Perclose
  • Knu-Knit, Dermabond, Lido with EPI
  • 0.035 x 260 J-tip wire
  • 6fr radial sheath (used for the pigtail for pressure monitoring)
  • 6Fr Balloon wedge PA catheter
  • VersaCross transeptal system
  • Long taper dilators 16fr, 24fr for access site dilation in preparation for Guide introduction
  • 24fr DrySeal
  • 10mm Mustang for septostomy or a 14mm Armada
  • (2) 5fr Pigtail catheters or 1 double lumen pigtail catheter (Langston)
  • RV Pacing setup - There is no wire in the LV that could be used for pacing
  • Stand support plate (example: plate used for MitraClip) placed under patient’s legs
  • 2 transducer sets (RHC; Ao/LV monitoring)
  • Implant and Delivery System Preparation -Back table
  • (3) Sterile Bowls (1000ml or larger)
  • 4L heparinized saline
  • (4) Luer lock syringes (50ml or larger)
  • (1) Pickup
  • (1) Y-Tuohy Borst Sidearm Connector
  • (1) 3-way Stopcock
  • (1) Pressure bag
  • (3) 6ft Sterile Hemostatic lines
  • IV pole on wheels
  • Extra sterile gowns and gloves available for 4C team

Bail Out Equipment List

(List provided by Vendor)

  • ECMO
  • IABP
  • Impella CP
  • Blood transfusion ability
  • Pacemaker interrogator if applicable per patient (for rapid pacing)
  • Peripheral vascular balloons – over the wire, 0.035” wire compatible Septal Closure and PVL Instrument List
  • Septal Closure Devices available – 2 per size: • Gore Cardioform PFO: 20mm, 25mm, 30mm • Abbott Amplatzer Cribaform: 18mm, 25mm, 30mm, 35mm • Corresponding delivery systems to match the above devices
  • PVL plug devices available - (for example: Amplatzer ADO2 in these sizes: 4x6mm, 5x4mm, 5x6mm, 6x4mm, 6x6mm) Corresponding delivery systems to match the above devices (must be at least 80cm in length)
  • 110cm Cook Ansel or Shuttle in both 4 French and 5 French