Quantien RFR wire (Abbott): Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
No edit summary |
||
(One intermediate revision by the same user not shown) | |||
Line 4: | Line 4: | ||
This is the iFR and FFR wire currently used in the cath lab. | This is the iFR and FFR wire currently used in the cath lab. | ||
Both console and wire transmitter talk to the "Wi-Box" mounted on the side of the table. The pressure manifold being used must be plugged into this box, and the output from the box to the GE box will be the AO pressure used for the measurement. | |||
See details on the [[Pressure Wiring Diagrams]] page | |||
*Unit will generally be kept behind Room 6 | *Unit will generally be kept behind Room 6 | ||
Latest revision as of 19:09, 5 November 2024
This is the iFR and FFR wire currently used in the cath lab.
Both console and wire transmitter talk to the "Wi-Box" mounted on the side of the table. The pressure manifold being used must be plugged into this box, and the output from the box to the GE box will be the AO pressure used for the measurement.
See details on the Pressure Wiring Diagrams page
- Unit will generally be kept behind Room 6
Procedure
- Plug console in and turn on.
- Log into the device username:cpmcvn, password:cmpcvn
- If device was powered off less than 5 minutes ago, it will ask if you want to continue the same patient.
- Select the room you are in.
- Enter patient info. (Can be entered at the end for STEMI)
- Follow on-screen prompts
- Re-zero PA pressure on device and (Optional)McKesson
- Flush wire and place flat on table
- Press connect on machine, then turn on wire transmitter
- Light should be solid green if connected
- Introduce wire into guide and advance until radio-lucent portion is outside the guide
- Equalize
Note: RFR may not show a steady 1.0 due to the way it is calculated
- Select mode at top, center of screen
- RFR - Resting full-cycle Ratio (AKA - IFR)
- Average of 5-beat difference through full cardiac cycle
- Camera icon gives a single reading
- Pullback button activates pullback
- FFR - Traditional measurement using Adenosine
- Follow physicians guidance for Adenosine and recording
- Unit will display lowest measurement from recording
- RFR - Resting full-cycle Ratio (AKA - IFR)
- Bottom left of screen contains options for vessel and drug labels
- Buttons at bottom of screen allow for switching between review and live screens
- Unit is not set up for archive as of yet. Record results in McKesson
- Unplug unit to turn off. This is OK per the rep.
Troubleshooting
- Turn everything off (Unplug)
- Turn recording console back on, and re-pair with wireless transmitter