Septostomy: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
|||
(2 intermediate revisions by one other user not shown) | |||
Line 8: | Line 8: | ||
* Indiflator | * Indiflator | ||
* Naked 3 port manifold for LA pressure | * Naked 3 port manifold for LA pressure | ||
* Sheath , one French size larger then minimum balloon size | |||
* Balloon, size determined via ECHO | |||
===* NEW - Baylis Transseptal=== | ===* NEW - Baylis Transseptal=== | ||
Line 23: | Line 25: | ||
== Equipment: Spies == | == Equipment: Spies == | ||
* | * Indeflator | ||
* Naked 3 port manifold for LA pressure | * Naked 3 port manifold for LA pressure | ||
===* NEW - Baylis Transseptal=== | ===* NEW - Baylis Transseptal=== | ||
* VersaCross system | * VersaCross system | ||
Line 37: | Line 36: | ||
*.032 X 260 “J” wire | *.032 X 260 “J” wire | ||
*SuperStiff 1cm Tip 260cm | |||
*5F MPA2 | |||
===Procedure requires Baylis=== | ===Procedure requires Baylis=== | ||
*Baylis generator | *Baylis generator | ||
Line 43: | Line 44: | ||
===Balloon size determined by MD.=== | ===Balloon size determined by MD.=== | ||
*Sheath size determined by balloon. Balloon volume listed on Structural Heart Cart, and chart above. | *Sheath size determined by balloon. Balloon volume listed on Structural Heart Cart, and chart above. | ||
*Have a 22 | *Have a 22 True balloon on standby for Spies and Daniels | ||
<br /> | <br /> | ||
<hr /> | <hr /> | ||
Approved: MD initials MM/YY | Approved: MD initials MM/YY | ||
[[Category:Procedures]] | [[Category:Procedures]] |
Latest revision as of 19:40, 11 October 2024
An atrial Septostomy is performed in the setting of ECMO when there is a concern that Left Atrial pressures are high due to the ECMO flow. Generally done in the setting of ECMO. It can be a stand alone procedure if not performed when ECMO is initiated.
Equipment: Daniels
- Indiflator
- Naked 3 port manifold for LA pressure
- Sheath , one French size larger then minimum balloon size
- Balloon, size determined via ECHO
* NEW - Baylis Transseptal
- VersaCross system
- Grounding pad
STANDBY!
- SL1 TransSeptal sheath
- Large curve Baylis needle
- Baylis cable
- .032 X 260 “J” wire
- Amplatz SuperStiff 1cm Tip 260cm
- 5F MPA2
Equipment: Spies
- Indeflator
- Naked 3 port manifold for LA pressure
* NEW - Baylis Transseptal
- VersaCross system
- Grounding pad
STANDBY!
- SL1 TransSeptal sheath
- Large curve Baylis needle
- Baylis cable
- .032 X 260 “J” wire
- SuperStiff 1cm Tip 260cm
- 5F MPA2
Procedure requires Baylis
- Baylis generator
- Grounding Pad
Balloon size determined by MD.
- Sheath size determined by balloon. Balloon volume listed on Structural Heart Cart, and chart above.
- Have a 22 True balloon on standby for Spies and Daniels
Approved: MD initials MM/YY