Aveir: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
No edit summary
Line 17: Line 17:
-4x pressure tubing (labeled A, B, C, D)
-4x pressure tubing (labeled A, B, C, D)


-50-60cc syringe
-50-60cc syringe x2


-micropuncture + 8fr pinnacle (or 8f dilator)
-8fr pinnacle (or 8f dilator)


-7cm floppy tip 180cm amplatz super stiff
-7cm floppy tip 180cm .035" amplatz super stiff


-dilators 12/18/24 short cooks
-dilator 20f coons


-0.0 PDS + needle driver
-0.0 PDS + needle driver
-radpad
-rampart cover
-ioban
-access cover





Revision as of 22:27, 16 January 2024

RN:

-2x 1L bags NS with 2000u heparin per liter bag. One bag on gravity (these 2 lines will be labeled A/B), one on pressure (these 2 lines will be labeled C/D). 4x non pump tubings (2 per bag) labeled as listed above, spiked and primed. Optionally, if you want to clear the air from the gravity (A/B) lines faster, you can put it on pressure to flush, then immediately come off pressure after they're flushed, as they need to be on gravity during the case per protocol.

-double label each line - one on roller clamp and one on the end where you attach to the extension.

-MD will verbalize when to flush and when to set to 1 drop per second (via roller clamp.)

-1L heparinized saline (10,000u heparin) on the scrub table

-50cc omnipaque or visipaque if indicated (can be cut to ~50/50)

-lidocaine/bupivicaine

Tech:

-4x pressure tubing (labeled A, B, C, D)

-50-60cc syringe x2

-8fr pinnacle (or 8f dilator)

-7cm floppy tip 180cm .035" amplatz super stiff

-dilator 20f coons

-0.0 PDS + needle driver

-radpad

-rampart cover

-ioban

-access cover




created 03/29/23 zr