Septostomy: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
No edit summary
Line 5: Line 5:
An atrial Septostomy is performed in the setting of ECMO when there is a concern that Left Atrial pressures are high due to the ECMO flow. Generally done in the setting of ECMO. It can be a stand alone procedure if not performed when ECMO is initiated.
An atrial Septostomy is performed in the setting of ECMO when there is a concern that Left Atrial pressures are high due to the ECMO flow. Generally done in the setting of ECMO. It can be a stand alone procedure if not performed when ECMO is initiated.


===Equipment:===
== Equipment: Daniels ==
 
* Indiflator
* Indiflator
* Naked 3 port manifold for LA pressure
* Naked 3 port manifold for LA pressure
Line 22: Line 21:


*5F MPA2
*5F MPA2
== Equipment: Spies ==
* Indiflator
* Naked 3 port manifold for LA pressure
* Amplatz SuperStiff 1cm Tip 260cm
*5F MPA2
===* NEW - Baylis Transseptal===
* VersaCross system
*Grounding pad
===STANDBY!===
*SL1 TransSeptal sheath
*Large curve Baylis needle
*Baylis cable
*.032 X 260 “J” wire
===Procedure requires Baylis===
===Procedure requires Baylis===
*Baylis generator
*Baylis generator

Revision as of 21:32, 4 January 2024

Z-Med Inflation Chart
Other Balloons Inflation Chart

An atrial Septostomy is performed in the setting of ECMO when there is a concern that Left Atrial pressures are high due to the ECMO flow. Generally done in the setting of ECMO. It can be a stand alone procedure if not performed when ECMO is initiated.

Equipment: Daniels

  • Indiflator
  • Naked 3 port manifold for LA pressure

* NEW - Baylis Transseptal

  • VersaCross system
  • Grounding pad

STANDBY!

  • SL1 TransSeptal sheath
  • Large curve Baylis needle
  • Baylis cable
  • .032 X 260 “J” wire
  • Amplatz SuperStiff 1cm Tip 260cm
  • 5F MPA2

Equipment: Spies

  • Indiflator
  • Naked 3 port manifold for LA pressure
  • Amplatz SuperStiff 1cm Tip 260cm
  • 5F MPA2

* NEW - Baylis Transseptal

  • VersaCross system
  • Grounding pad

STANDBY!

  • SL1 TransSeptal sheath
  • Large curve Baylis needle
  • Baylis cable
  • .032 X 260 “J” wire

Procedure requires Baylis

  • Baylis generator
  • Grounding Pad

Balloon size determined by MD.

  • Sheath size determined by balloon. Balloon volume listed on Structural Heart Cart, and chart above.
  • Have a 22 Tyshak nucleus on standby for Spies and Daniels



Approved: MD initials MM/YY