BASILICA: Difference between revisions

From Bay Area Structural Heart Wiki
(added procedure steps)
No edit summary
 
(21 intermediate revisions by 4 users not shown)
Line 5: Line 5:
===Open===
===Open===


*14frx33cm Gore Dry Seal Flex
*14 F Dryseal sheath for Sapien or for in-line Evolute ; 18Fr for non-inline Evolute (Confirm before opening)
*Amplatz Extra Stiff for Sheath advancement
*Suture for dryseal securement
*Micropuncture
*7F sheath
*8F sheath
*5/6 Slender (2.5 of Verapamil)
*Grandslam 300
*(2) .035  x 150 J wires
*.035 x 260 J wire
*Terumo Stiff Straight Glidewire 0.035 x 150 cm
*Safari Wire
*(3) Perclose
*.018 Platinum Plus guidewire 260 cm or any other .018 wire
*Gooseneck Snare 20mm- '''VERIFY SIZE BEFORE OPENING'''
*6fr MP1 guide catheter
*6fr MP1 guide catheter
*.018 Platinum Plus guidewire 260 cm
*7 Fr AL3 (for LCA BASILICA)  -OR- 7F MP1 Guide (For RCA BASILICA)
*Gooseneck Snare 20mm
*5Fr AL1 diagnostic
*6 Fr PAL 1.0 (for LCA BASILICA)  -OR- 6 FR MPA1 Guide (For RCA BASILICA)
*(2) 5Fr Ang Pigtail
*.014 Astato xs 20 guidewire 300 cm
*6Fr JL4/JR4 (when doing coronary angiography)
 
*.014 Astato xs 20 guidewire 180cm
*Piggyback wire converter (structural cart) -OR- 135 FineCross
*Bovie pen
*Bovie pen
**Grounding Pad
**Grounding Pad
Line 17: Line 32:
*(2) Watchdog devices
*(2) Watchdog devices
*(2) Stopcocks
*(2) Stopcocks
*4 hemostats
*(4) Needle drivers
*(3) 60ml syringes
*(2) 12 inch extension tubing
*Alligator pacing wire
*Sentinel (don't open until we pass the Grandslam)


===Standby===
===Standby - PULLED and HANGING on WALL===


*Tip deflecting wire
*5Fr 125 IMA
*5Fr 125 MP1
*6Fr Guideliner
=== '''Additional Standby Items - don't need to have pulled''' ===
*Gooseneck snares 25, 30mm
*Gooseneck snares 25, 30mm
*Piggyback wire converter 4fr x 145 (On standby)
*.035 QuickCross
*.035 QuickCross (On standby)
*AL2, AL3, AL4,  EBU4, EBU 3.75 (Left leaflet)
*AL2, AL3, EBU4, EBU 3.75 (Left leaflet)
*MPA1, JR4, IM, AR1 (Right leaflet)
*PAR1, JR4, IM, AR1 (Right leaflet)
*(3) .014 torque devices
*(2) .035 torque devices
*Finecross 135


<br />
<br />


=== Procedure ===
===Procedure===
 
*Setup and access is the same as TAVR '''''except''''' a 14F Dryseal sheath is inserted initially instead of the ESheath.
**Bovie set at 30 CUT ONLY.
*AL1 advanced over J wire through 7F sheath.  Wire exchanged for straight stiff glidewire for valve crossing.
*AL1 exchanged over the wire for MP1 Guide with watchdog and stopcock attached.
*Platinum Plus wire is preshaped by MD and inserted through MP to retain LV access.
*Snare advanced through MP alongside Platinum Plus.
*BASILICA catheter advanced through Dryseal sheath with watchdog and stopcock attached.
*Kinked Astato wire advanced through BASILICA catheter.
*Hand injection performed to verify position of BASILICA catheter.
*Astato wire connected to bovie with 2 hemostats.
*RF applied as Astato wire advanced.
*Once wire penetrates leaflet it is snared and pulled into MP1.
*Platinum Plus wire removed
*AL1 advanced next to BASILICA catheter and valve recrossed in standard fashion.
*60ml D5 filled syringes and extension tubing attached to sideports of guides and injected simultaneously during leaflet laceration.
*MP removed. BASILICA catheter removed.
*Amplatz ES advanced through AL1 and AL1 removed.
*Pigtail advanced to coronary cusp
*Dryseal exchanged for ESheath
*Do a TAVR.
 
== Double BASILICA ==
 
* 18F Dryseal and 14F Dryseal (No 7F sheath)
* (2) 6F MP1
* 7 Fr AL2 or 3? (for LCA BASILICA)  -AND- 7F MP1 Guide (For RCA BASILICA)
* (2) Goose Neck Snares (one platinum plus, once first leaflet is traversed the platinum plus is removed and can be used for other leaflet)
* (4) Watchdogs and stopcocks
* (6) Needle drivers
* (2) .014 Astato xs 20 guidewire 180cm
*(2) Piggyback wire converter 4fr x 145
 
18F Dryseal on primary side. Once both leaflets are traversed, do valve check. Platinum Plus removed. AL1 advanced and valve recrossed. Pigtail left in LV for retention. Right leaflet is lacerated, equipment removed. Pigtail removed over delivery wire. Valve delivery advanced and left indwelling in aorta. Left leaflet lacerated and equipment removed. Valve deployed.


* Setup and access is the same as TAVR '''''except''''' a 14F Dryseal sheath is inserted initially instead of the ESheath.
** Bovie set at 30 CUT ONLY.
* AL1 advanced over J wire through 6F sheath.  Wire exchanged for straight stiff glidewire for valve crossing.
* AL1 exchanged over the wire for MP1 Guide with watchdog and stopcock attached.
* Platinum Plus wire is preshaped by MD and inserted through MP to retain LV access.
* Snare advanced through MP alongside Platinum Plus.
* BASILICA catheter advanced through Dryseal sheath with watchdog and stopcock attached.
* Kinked Astato wire advanced through BASILICA catheter.
* Hand injection performed to verify position of BASILICA catheter.
* Astato wire connected to bovie with 2 hemostats.
* RF applied as Astato wire advanced.
* Once wire penetrates leaflet it is snared and pulled into MP1. 
* Platinum Plus wire removed
* Pacing wire advanced and tested
* Al1 advanced next to BASILICA catheter and valve recrossed in standard fashion.
* Leaflet lacerated
* MP removed. BASILICA catheter removed.
* Amplatz ES advanced through AL1 and AL1 removed.
* Pigtail advanced to coronary cusp
* Dryseal exchanged for ESheath
* Do a TAVR.





Latest revision as of 23:53, 3 January 2024

(Bioprosthetic Aortic Scallop Intentional Laceration to prevent Iatrogenic Coronary Artery obstruction)

ACC.org BASILICA Page

Open

  • 14 F Dryseal sheath for Sapien or for in-line Evolute ; 18Fr for non-inline Evolute (Confirm before opening)
  • Amplatz Extra Stiff for Sheath advancement
  • Suture for dryseal securement
  • Micropuncture
  • 7F sheath
  • 8F sheath
  • 5/6 Slender (2.5 of Verapamil)
  • Grandslam 300
  • (2) .035 x 150 J wires
  • .035 x 260 J wire
  • Terumo Stiff Straight Glidewire 0.035 x 150 cm
  • Safari Wire
  • (3) Perclose
  • .018 Platinum Plus guidewire 260 cm or any other .018 wire
  • Gooseneck Snare 20mm- VERIFY SIZE BEFORE OPENING
  • 6fr MP1 guide catheter
  • 7 Fr AL3 (for LCA BASILICA) -OR- 7F MP1 Guide (For RCA BASILICA)
  • 5Fr AL1 diagnostic
  • (2) 5Fr Ang Pigtail
  • 6Fr JL4/JR4 (when doing coronary angiography)
  • .014 Astato xs 20 guidewire 180cm
  • Piggyback wire converter (structural cart) -OR- 135 FineCross
  • Bovie pen
    • Grounding Pad
    • Bovie machine
  • (2) Watchdog devices
  • (2) Stopcocks
  • (4) Needle drivers
  • (3) 60ml syringes
  • (2) 12 inch extension tubing
  • Alligator pacing wire
  • Sentinel (don't open until we pass the Grandslam)

Standby - PULLED and HANGING on WALL

  • Tip deflecting wire
  • 5Fr 125 IMA
  • 5Fr 125 MP1
  • 6Fr Guideliner

Additional Standby Items - don't need to have pulled

  • Gooseneck snares 25, 30mm
  • .035 QuickCross
  • AL2, AL3, AL4, EBU4, EBU 3.75 (Left leaflet)
  • MPA1, JR4, IM, AR1 (Right leaflet)


Procedure

  • Setup and access is the same as TAVR except a 14F Dryseal sheath is inserted initially instead of the ESheath.
    • Bovie set at 30 CUT ONLY.
  • AL1 advanced over J wire through 7F sheath. Wire exchanged for straight stiff glidewire for valve crossing.
  • AL1 exchanged over the wire for MP1 Guide with watchdog and stopcock attached.
  • Platinum Plus wire is preshaped by MD and inserted through MP to retain LV access.
  • Snare advanced through MP alongside Platinum Plus.
  • BASILICA catheter advanced through Dryseal sheath with watchdog and stopcock attached.
  • Kinked Astato wire advanced through BASILICA catheter.
  • Hand injection performed to verify position of BASILICA catheter.
  • Astato wire connected to bovie with 2 hemostats.
  • RF applied as Astato wire advanced.
  • Once wire penetrates leaflet it is snared and pulled into MP1.
  • Platinum Plus wire removed
  • AL1 advanced next to BASILICA catheter and valve recrossed in standard fashion.
  • 60ml D5 filled syringes and extension tubing attached to sideports of guides and injected simultaneously during leaflet laceration.
  • MP removed. BASILICA catheter removed.
  • Amplatz ES advanced through AL1 and AL1 removed.
  • Pigtail advanced to coronary cusp
  • Dryseal exchanged for ESheath
  • Do a TAVR.

Double BASILICA

  • 18F Dryseal and 14F Dryseal (No 7F sheath)
  • (2) 6F MP1
  • 7 Fr AL2 or 3? (for LCA BASILICA) -AND- 7F MP1 Guide (For RCA BASILICA)
  • (2) Goose Neck Snares (one platinum plus, once first leaflet is traversed the platinum plus is removed and can be used for other leaflet)
  • (4) Watchdogs and stopcocks
  • (6) Needle drivers
  • (2) .014 Astato xs 20 guidewire 180cm
  • (2) Piggyback wire converter 4fr x 145

18F Dryseal on primary side. Once both leaflets are traversed, do valve check. Platinum Plus removed. AL1 advanced and valve recrossed. Pigtail left in LV for retention. Right leaflet is lacerated, equipment removed. Pigtail removed over delivery wire. Valve delivery advanced and left indwelling in aorta. Left leaflet lacerated and equipment removed. Valve deployed.





Approved: MM/YY