In the course of a TAVI, it is sometimes necessary to implant a leadless pacemaker. This facility uses the Micra device. This page will guide you on this procedure in the post-TAVI setting.
In the course of a TAVI, it is sometimes necessary to implant a leadless pacemaker. This facility uses the Micra device. This page will guide you on this procedure in the post-TAVI setting.
== MORE INFO LOCATED ON MICRA-ROMICK PAGE ==


===Already on the table:===
===Already on the table:===
*Micropuncture set
*Micropuncture set
*150 j-wire
*150 J-wire
*2x 20cc syringes
*2x 20cc syringes
*1-2x 60cc syringes
*1-2x 60cc syringes


===Open:===
===Open:===
*14 and 22 coons dilators
*12/18/24 dilators
*0 PDS or Prolene
*0 PDS  
*180 Amplatz Super Stiff, 7cm tip
*180 Amplatz Super Stiff, 7cm tip
*Micra sheath from behind EP


If pt. is dependent on TPM and it is in the RFV, assume you will be switching the TPM to the LFV and open:
If pt. is dependent on TPM and it is in the RFV, assume you will be switching the TPM to the LFV and open:
Line 17: Line 20:
*5fr. Ballon tip TPM catheter
*5fr. Ballon tip TPM catheter


Prep:
===Prep:===
*Set ACIST at 5ml/sec, 20ml total (to fill syringes).
*Set ACIST at 5ml/sec, 20ml total (to fill syringes).
*
*20 and 60cc syringes filled with Hep Saline
*Flush and assemble Micra Sheath
*RN preps hep saline on IV pump @ 125ml/hr
 
===Procedure:===
====If on TPM====
#Access LFV and insert 8.5 Brite Tip
#Insert balloon-tip TPM and establish capture
#Wire RFV with Amplatz SS
#Dilate and place Micra sheath
#Attach Hep Saline drip
#Micra delivery system is flushed gently with 60cc syringe and inserted
#Fill delivery system with contrast
#Before deployment, do LAO with contrast
#Do RAO with saline flush (clears delivery system)
#Deploy Micra
#Tug test under 30 fps. cine
#Flush delivery system with full 60 cc syringe
#Pull suture
#Remove delivery system
#Place purse-string and woggle
#Remove sheath
 
[[Category:Procedures]]

Latest revision as of 22:12, 5 April 2023

In the course of a TAVI, it is sometimes necessary to implant a leadless pacemaker. This facility uses the Micra device. This page will guide you on this procedure in the post-TAVI setting.

MORE INFO LOCATED ON MICRA-ROMICK PAGE

Already on the table:

  • Micropuncture set
  • 150 J-wire
  • 2x 20cc syringes
  • 1-2x 60cc syringes

Open:

  • 12/18/24 dilators
  • 0 PDS
  • 180 Amplatz Super Stiff, 7cm tip
  • Micra sheath from behind EP

If pt. is dependent on TPM and it is in the RFV, assume you will be switching the TPM to the LFV and open:

  • New ultrasound probe cover
  • 8.5 x 35 Brite-tip sheath
  • 5fr. Ballon tip TPM catheter

Prep:

  • Set ACIST at 5ml/sec, 20ml total (to fill syringes).
  • 20 and 60cc syringes filled with Hep Saline
  • Flush and assemble Micra Sheath
  • RN preps hep saline on IV pump @ 125ml/hr

Procedure:

If on TPM

  1. Access LFV and insert 8.5 Brite Tip
  2. Insert balloon-tip TPM and establish capture
  3. Wire RFV with Amplatz SS
  4. Dilate and place Micra sheath
  5. Attach Hep Saline drip
  6. Micra delivery system is flushed gently with 60cc syringe and inserted
  7. Fill delivery system with contrast
  8. Before deployment, do LAO with contrast
  9. Do RAO with saline flush (clears delivery system)
  10. Deploy Micra
  11. Tug test under 30 fps. cine
  12. Flush delivery system with full 60 cc syringe
  13. Pull suture
  14. Remove delivery system
  15. Place purse-string and woggle
  16. Remove sheath