Aveir: Difference between revisions

From Bay Area Structural Heart Wiki
(Created page with "RN: -2x 1L bags NS with 2000u heparin per liter bag. One bag on gravity (these 2 lines will be labeled A/B), one on pressure (these 2 lines will be labeled C/D). 4x non pump tubings (2 per bag) labeled as listed above, spiked and primed. Optionally, if you want to clear the air from the gravity (A/B) lines faster, you can put it on pressure to flush, then immediately come off pressure after they're flushed, as gravity is required per protocol. -double label each line -...")
 
No edit summary
Line 1: Line 1:
RN:
'''<big>RN:</big>'''
 
-2x 1L bags NS with 2000u heparin per liter bag. One bag on gravity (these 2 lines will be labeled A/B), one on pressure (these 2 lines will be labeled C/D). 4x non pump tubings (2 per bag) labeled as listed above, spiked and primed. Optionally, if you want to clear the air from the gravity (A/B) lines faster, you can put it on pressure to flush, then immediately come off pressure after they're flushed, as gravity is required per protocol.  
-2x 1L bags NS with 2000u heparin per liter bag. One bag on gravity (these 2 lines will be labeled A/B), one on pressure (these 2 lines will be labeled C/D). 4x non pump tubings (2 per bag) labeled as listed above, spiked and primed. Optionally, if you want to clear the air from the gravity (A/B) lines faster, you can put it on pressure to flush, then immediately come off pressure after they're flushed, as gravity is required per protocol.  


Line 12: Line 13:
-lido/bupi
-lido/bupi


Tech:
'''<big>Tech:</big>'''
 
-4x pressure tubing (labeled A, B, C, D)
-4x pressure tubing (labeled A, B, C, D)


Line 24: Line 26:


-0.0 PDS + needle driver
-0.0 PDS + needle driver
created 3/29 zr

Revision as of 19:30, 29 March 2023

RN:

-2x 1L bags NS with 2000u heparin per liter bag. One bag on gravity (these 2 lines will be labeled A/B), one on pressure (these 2 lines will be labeled C/D). 4x non pump tubings (2 per bag) labeled as listed above, spiked and primed. Optionally, if you want to clear the air from the gravity (A/B) lines faster, you can put it on pressure to flush, then immediately come off pressure after they're flushed, as gravity is required per protocol.

-double label each line - one on roller clamp and one on the end where you attach to the extension.

-MD will verbalize when to flush and when to set to 1 drop per second (via roller clamp.)

-1L heparinized saline on the scrub table

-50cc omnipaque or visi if indicated (can be cut to ~50/50)

-lido/bupi

Tech:

-4x pressure tubing (labeled A, B, C, D)

-50-60cc syringe

-micropuncture + 8fr pinnacle (or 8f dilator)

-7cm floppytip 180cm amplatz xtra stiff

-dilators 12/18/24 short cooks

-0.0 PDS + needle driver



created 3/29 zr